WPI IGS100 Instruction Manual

WPI IGS100 Instruction Manual

Implantable glucose sensor

Advertisement

Quick Links

IGS100
Implantable Glucose Sensor
INSTRUCTION MANUAL
Serial No._____________________
011310
World Precision Instruments

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the IGS100 and is the answer not in the manual?

Questions and answers

Summary of Contents for WPI IGS100

  • Page 1 IGS100 Implantable Glucose Sensor INSTRUCTION MANUAL Serial No._____________________ 011310 World Precision Instruments...
  • Page 2 IGS100 World Precision Instruments...
  • Page 3: Table Of Contents

    Parts list ..........................2 unpacking ........................2 oPerAtINg INStructIoNS ...................3 conditioning the Sensor ....................3 Preparing Solution for calibration ................3 calibrating the IgS100 ....................3 SeNSor cAre ANd MAINteNANce ................5 Storing the Sensor ......................5 Maintaining the Sensor ....................5 Sterilizing the Sensor ....................5 AcceSSorIeS ........................6 troubleShootINg ......................6...
  • Page 4 IGS100 World Precision Instruments...
  • Page 5: About This Manual

    Please read these carefully. NoteS and tIPS contain helpful information. Fig. 1–IGS100 sensor Fig. 2–Tip of the sensor showing the coil holding the enzyme saturated cotton plug. The diameter of the coiled tip is about 600µm.
  • Page 6: Introduction

    IGS100 may be used in vivo for weeks or even months. When not in use, it has a six-month shelf life.
  • Page 7: Operating Instructions

    IGS100 Returns: do not return any goods to WPI without obtaining prior approval (rMA # required) and instructions from WPI’s returns department. goods returned (unauthorized) by collect freight may be refused. If a return shipment is necessary, use the original container, if possible. If the original container is not available, use a suitable substitute that is rigid and of adequate size.
  • Page 8 IGS100 temperature, and allowing the temperature of the solution in the bottle to equilibrate with water bath. 5. once the current has achieved a stable value, record this value or use zero function in the instrument to set the baseline to zero.
  • Page 9: Sensor Care And Maintenance

    Maintaining the Sensor After each use, rinse the sensor with distilled water. Sterilizing the Sensor the sensor can be sterilized chemically using an appropriate disinfectant (for example, cidex, WPI#7364). CAUTION: do not use alcohol on the sensor. World Precision Instruments...
  • Page 10: Accessories

    Part Number Description 91580 Microsensor cable TROUbLeShOOTING If you have a problem/issue with your IGS100 sensor, contact the WPI technical Support team at 941.371.1003 or technicalsupport@wpiinc.com. SpeCIfICATIONS the IGS100 sensor conforms to the following specifications: coefficient of variation (cV) ≤5%...
  • Page 11: Warranty

    WPI shall not be liable for any damage to data or property that may be caused directly or indirectly by use of this product.
  • Page 12 International Trade Center, 175 Sarasota Center Blvd., Sarasota FL 34240-9258 Tel: 941-371-1003 • Fax: 941-377-5428 • E-mail: sales@wpiinc.com Astonbury Farm Business Centre • Aston, Stevenage, Hertfordshire SG2 7EG Tel: 01438-880025 • Fax: 01438-880026 • E-mail: wpiuk@wpi-europe.com Germany Zossener Str. 55, 10961 Berlin Tel: 030-6188845 •...