Page 1
WORLD PRECISION INSTRUMENTS INSTRUCTION MANUAL ISO-HPO-2 Hydrogen Peroxide Sensor for use with WPI Analyzers Serial No._____________________ www.wpiinc.com 031919...
(#600012) and an internal refillable electrolyte (#100042). The sensor is designed for use in cell cultures and similar applications. The basic structure of the ISO-HPO-2 sensor is shown in Fig. 2. It consists of an internal H -sensing working/reference electrode combination. This electrode fits...
Concealed damage should be reported at once to the carrier and an inspection requested. Please read the section entitled “Claims and Returns” on page 11 of this manual. Please contact WPI Customer Service if any parts are missing at 941.371.1003 or customerservice@wpiinc.com.
(included). See “Replacing the Membrane Sleeve” on page 7. Immerse the ISO-HPO-2 sensor tip in PBS buffer solution. The sensor tip should be immersed about 0.3-0.5mm into the solution, and should not be touched by the stir bar.
0.1M PBS buffer solution (PBS buffer tablets) (To make standard solution, see “Making of 0.1M PBS Buffer Solution” on page 5.) NOTE: The multi-port measurement chamber (WPI #NOCHM-4) can be used as an alternative calibration kit, specifically for use at different temperature condition.
ISO-HPO-2 Add 231mL of 2-4% H (Sigma-Aldrich 323381) to the flask. Then, use the removed 1mL of solution from step 2 to dilute the solution in the flask to exactly 250mL again. Store this standard solution in an amber bottle, if available, or alternatively, wrap aluminum foil around the bottle to limit light intrusion.
From this output, a calibration curve Fig. 4 can be created by plotting the changes in current (pA) against the changes in concentration (mM). The slope of this curve indicates sensitivity of the probe. Fig. 4—Calibration Data Interference Temperature The background current of the sensor will usually increase with increasing temperature of the experiment.
Storing the Sensor When the ISO-HPO-2 sensor is not being used (for short-term storage) it should be connected to the TBR4100/1025 with power ON and with the tip suspended in 0.1M PBS solution. (Do not use water.) This will keep the sensor polarized and ready for immediate use.
Page 12
Additional membrane sleeves (WPI#600012) may be pur- chased separately. Dip the internal electrode 1–2cm into the ISO-HPO electrolyte (WPI #100042) provided in the start-up kit. The current will rise rapidly offscale. Using the Mi- croFil™ nonmetallic syringe needle (WPI #MF28G67-5) and 1mL plastic syringe (included in the Startup kit) inject approximately 100µL of electrolyte directly into...
(1M). If the current increases dramatically or is offscale then the membrane integrity is not good and a new membrane will have to be fitted. Additional ISO-HPO-2 membranes (packages of 4) are available (WPI #600012).
See “Replacing the Membrane Sleeve” on evaporated. page 7. NOTE: If you have a problem/issue that falls outside the definitions of this troubleshooting section, contact the WPI Technical Support team at (941) 371-1003 or technicalsupport@wpiinc.com. SPECIFICATIONS The sensor conforms to the following specifications: Outside Diameter ........................2 mm...