Summary of Contents for Nippon Genetics Fast Gene NG020A

  • Page 1 Mini Dry Bath Advance NG020A Manual NIPPON Genetics EUROPE GmbH Mariaweilerstraße 28-30, 52349 Düren, Germany, www.nippongenetics.eu Tel: +49 2421 554960 Fax: +49 2421 55496 11, E-Mail: info@nippongenetics.de,...
  • Page 2 FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 3: Table Of Contents

    3.5.2 Calibration Procedure..........P. 17 P. 18 3.6 Automatic Heating..............3.7 System Maintenance..............P. 19 3.8 Troubleshooting................. P. 19 P. 20 4. ORDERING INFORMATION..............5. WARRANTY....................P. 21 5. CONTACT INFORMATION..............P. 21 FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 4: General Information

    Please check the instrument and the packing list when you first open the packing case. If you find there is something wrong with the instrument or the packaging content, please contact Nippon Genetics Europe or your local distributor. Packing List:...
  • Page 5: Safety

    • Do not use the unit in the presence of flammable or combustible materials, as it may result in fire or explosion. This device contains components which may ignite such materials. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 6: Safety Information

    • 75 kPa – 106 kPa • Altitude must not exceed 2,000 meters • Ambient to 40°C operating temperature • Mains supply voltage fluctuations up to ±10% of the normal voltage FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 7: Avoid Electrical Shock

    Follow the guidelines below to ensure a safe operation of the unit. The FastGene Mini Dry Bath Advance has been designed with shielded wires, ® thus minimizing any potential shock hazard to the user. Nippon Genetics EUROPE disadvises the use of unshielded wires. To avoid electrical Shock: •...
  • Page 8: Manual

    Max. temperature 100 °C Max. Power 50 W Cooling Time Natural Cooling Weight ≤ 1 kg Dimensions (H x D x W) 13.6 x 16 x 11 cm With Transparent Plastic Cover FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 9: Device Components

    3.2 Device Components Please refer to the following page for the location of controls and features: Front: Back: FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 10: Installation Guide

    After the device is turned on, it will automatically heat up to the set temperature for 5 min. Put the tubes into the the wells of the heating block and cover it with the transparent heating block cover. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 11: Operation Guide

    Start/Stop........Operation and control of the device Prog..........Program button, set the device (up and down)....Set program, temperature and time (left).......... Move left to choose parameter values (right)........Move right to choose parameter values FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 12: Program Setting

    To set the temperature of the specific program step, press the right button until the figures of the temperature start flashing and set the individual digits by pressing the up or down button (e.g set to 85 °C). FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 13 P3C 085.0 30 OFF In this example, the following step parameters have been set for program 3 (P3): P3A 037.0 20m P3B 085.0 30m P3C 085.0 30 OFF FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 14: Program Execution

    P3 has ended. The completion of the program is indicated by an alarm and displayed with "FINISHED" and the current heating block temperature: FINISHED 85.0 Press the Start/Stop button to return to the program selection page. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 15 Note: If the FastGene Mini Dry Bath Advance needs to be used at a very ® accurate temperature (± 0.3 °C), we recommend to preheat the thermo block for ~20 min. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 16: Temperature Calibration

    0.1 °C. The tip of the thermometer should be completely immerged in the oil. Use heat isolation material on the thermoblock to separate it from the surrounding (see figure below). FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 17: Calibration Procedure

    69.8 °C. Press the Start/Stop button to cofirm the correction. Repeat this procedure analogously for 100 °C. Return to the program selection page and if needed correct the temperature settings. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 18: Automatic Heating

    Press the program button (Prog) to save the current settings and return to the program selection page. Note: The default setting is "APF: ENABLE". With this setting, the system will automatically heat up after turning it on. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 19: System Maintenance

    Monitor displays Short circuit of "SHORT" with alarm Contact seller temeprature sensor sound Block not heating Broken heater Contact seller Button not working Broken button Contact seller FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 20: Ordering Information

    Block for NG028A Note: When using the thermoblocks NG027A for 6x 15 ml tubes and NG028A for 2x 50 ml tubes, the temperature needs to be set to a higher value. FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 21: Warranty

    5. WARRANTY Nippon Genetics EUROPE provides a product warranty to the purchaser against defects and/or incorrect material processing for one year after the shipping date. This warranty excludes damages resulting from shipping, misuse, carelessness, or neglect. Nippon Genetics EUROPE's liability under the warranty is limited to the receipt of adequate evidence by the customer that the defect falls under the warranty conditions.
  • Page 22 FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...
  • Page 23 FastGene Mini Dry Bath Advance ® NIPPON Genetics EUROPE...