Summary of Contents for Nippon Genetics FASTGENE ELECTRONIC PIPETTE-200

  • Page 2: Table Of Contents

    12-4 Weight calibration function (mg calibration function) ............30 13. ADJUSTING HEIGHT OF THE TIP EJECTOR ..................32 14. STORAGE AND MAINTENANCE ....................... 32 14-1 Autoclave ..........................33 15. TROUBLE SHOOTING ..........................34 16. WHEN REQUESTNG REPAIRE ......................... 35 17. SPECIFICATION ............................35 NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 3: For Safe Use

    - Should you get battery fluid from a leaking battery in your eye, immediately flush with copious amounts of water and seek immediate medical attention. Should the liquid contact the clothes or skin, rinse immediately with copious amounts of water. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 4: Introduction

    Impact-absorbing pads adopted to fully protect against falling. (Patent has been applied for)  *1 Calculated by operating force and movement *2 Approximately 1,800 dispensing operations are possible on a full charge NIPPON Genetics EUROPE Binsfelder Straße 77 52351 Düren Germany NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 5: Compliance

    Restriction of the use of certain Hazardous Substances The CE mark is an official mandatory European marking. Please note that any electronic product must comply with local laws and regulations when sold or used anywhere outside Europe. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 6: Fastgene ® Electronic Pipette - Function

    (Refer to “11-1 Volume calibration function (μL calibration function)”) (Patent applied for) Dispensing by weight is also available. Refer to “11-3 Dispensing in a unit of weight (in mg  unit)”) (Patent applied for) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 7: Packing Contents And Name Of Items

    ® (For FastGene electronic pipette -1200) Should the pipette arrive damaged or an accessory be missing, contact the nearest NIPPON Genetics EUROPE dealer. Note: The accessories included with this product may be changed without notice. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 8: Preparation Before Use

    Note: When connecting the battery to the pipette, all illuminations on the display illuminate and the pipette built-in piston automatically goes to the initial default position. If a key is pressed, the pipette goes into the operating mode. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 9: Recharging The Battery

    NOTE: After recharging the battery completely, remove the power cable from the pipette. Firmly attach the power connector cover by pushing it onto the pipette. Power connector cover Power connector Power cable (Mini USB B type) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 10: Exchange Selectable Power Plug

    Up key below the display. Operate the Operation key using the forefinger, as shown in the figure below. - Operate the eject button by using the thumb to remove the tip. Eject button Finger hook Operation key NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 11 Operation key Press and hold the Operation key on the pipette for at least five seconds until OFF is displayed on the display. Buzzer sounds (Three times), and the pipette turns the power off (OFF). NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 12: Parts Names

    Lower part of main Ejector body (Made of PP) Removable for autoclave (121℃ 2 atm 20 minutes) (Glass fiber is contained with 20%) Tip holder (Made of PVDF) (Front view) (Side view and rear view) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 13: Name And Functions Of Display And Keys

    Display COUNT:Shows the number of times the same operation is to number and be carried out. a number of PROG:Shows the stored number of user setting. (Refer to “9-5 time Program setting mode”.) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 14: Key Switches And Functions

    Operation key (key switch on rear side on the pipette) has the same function as the Up key. This allows you to quickly perform settings such as changing the volume without shifting the pipette in the hand. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 15: Function And How To Use

    [ 5 ] Press the Enter key to confirm dispensing amount. If you would like to change the dispensing amount, Press Enter key before aspirating. When setting, operate from step 4 as described above. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 16: Multiple Dispensing Mode (Md)

    1) Operation of the multiple dispensing mode This is a suitable function to dispense the same liquid continuously, such as when dispensing a liquid on a microplate, etc. The operation consists of aspirating one time and dispensing several times. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 17 Enter key to confirm dispensing amount for one time. [ 6 ] Press the Up or Down key to set dispensing count. * “Dispensing amount for one time x dispensing count” cannot be set if it exceeds the volume range. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 18 Operation key on the pipette to dispense the amount for one time. [ 3 ] In a similar manner, put the tip end into the next vessel and press the Operation key to dispense the liquid. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 19: Mixing Mode(Mix

    2) Selecting the mixing mode (The display is example.) [ 1 ] Press the Back key. [ 2 ] Press the Up or Down key to select “MIX”. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 20 3) Operating the mixing mode [ 1 ] Insert the tip end in the liquid to be mixed. [ 2 ] Press the Operation key on the pipette to aspirate the set mixing volume. Aspirating of mixed NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 21: System Setting Mode(Sys

    The SYS mode is used to perform or read out pipetting operation settings that suit the purpose or the liquid to be handled. 2) Operating the system setting mode (The display is example.) [ 1 ] Press the Back key. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 22 (Read out your preferred program from the programs previously set.) *3 The reverse operation is only selectable when blowout setting is off. It cannot be selected when the pipette is in MIX mode, either. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 23: Program Setting Mode

    To enable reverse operation, set the setting of at system setting mode (SYS). (Refer to “9-4 System setting mode (SYS)”.) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 24 [ 6 ] Turn the reverse operation mode ON using Up/Down key, then confirm this selection by pressing the Enter key. [ 7 ] End the system setting mode by pressing the Enter key while is being displayed. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 25: Blowout Function

    Operation key is being held down, and it returns to initial position when the finger was released from the Operation key. By releasing the Operation key after removing the tip end from the vessel, aspiration of the dispensed liquid in the tip again can be prevented. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 26: Total Discharge Function

    The reverse operation is recommended for a sample liquid that tends to linger in the tip. When dispensing the sample, touch the tip at a 30°to 40°angle to the wall of the receiving vessel. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 27: Calibrating The Pipette Using An Electronic Balance

    Selected dispensing volume setting for calibration (μL) (example is 100μL.) [ 1 ] Press the Back key. [ 2 ] Press and hold the Back key for approx. three (μL calibration mode) seconds while “AUTO” is selected. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 28: Resetting The Volume Calibration

    Go through the following procedure to restore the factory default settings for volume calibration: (The display is example.) [ 1 ] Press the Back key. [ 2 ] Press and hold the Back key for approx. three seconds while “AUTO” is selected. (μL calibration mode) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 29: Dispensing In A Unit Of Weight (In Mg Unit)

    * When the unit of weight (mg unit) is selected, perform weight calibration by the dispensing amount to be used. (The weight calibration data reverts back to the factory default once the unit is switched toμL.) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 30: Weight Calibration Function (Mg Calibration Function)

    The density of a liquid varies depending on the type and concentration of the material. Make sure to perform mg calibration when you dispense a different sample or use the mg unit for the first time. Further, when you need to always control dispensed amounts in a precise manner, perform mg NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 31 Up or Down key to alter the value to the actual measured weight. (The example is 95mg) [ 5 ] Press the Enter key to confirm the actual measured weight. (In this example, the dispensed amount is corrected and altered to 100mg.) NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 32: Adjusting Height Of The Tip Ejector

    Leakage within the pipette can be easily checked. - Pipette accuracy tester (Sold separately: BM series, AD-4212B-PT, FX-300i-PT): By using an electronic balance and specified software “WinCT-pipette”, the actual dispensed amount of the pipette can be confirmed. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 33: Autoclave

    - After performing autoclave, install the lower part of the pipette to the pipette after drying it completely. Keep the lower part of the pipette in a room for at least two hours before installing. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 34 Contact failure of the battery. Set the battery again. turned on. Dirt on the battery terminal. Replace the battery with a new one. Battery discharge. Recharge the battery. Deterioration of the battery. Replace the battery with a new one. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 35: Trouble Shooting

    ELECTRONIC PIPETTE -10 is ensured with 1.0μL or more. *2 When setting to standard mode, maximum aspirating and dispensing speed. *3 For recharging. The pipette can be used even when recharging Note: The specification may be changed without notice. NIPPON Genetics EUROPE – www.nippongenetics.eu...
  • Page 36 I attest to the fact that this pipette is free of contamination by any substances that could pose a health threat to humans, such as infectious bacteria or viruses, radioactive substances with associated risks of exposure, toxic substances, etc. Signature: Date: Company name (Facility name): Section name: Address: NIPPON Genetics EUROPE – www.nippongenetics.eu...

Table of Contents