Installation

Notes
Installation shall only be performed by qualified personnel and in accordance
with local govering regulations.
Do not mount the SITRANS LU SAM in direct sunlight without the use of a sun
shield.
This product is susceptible to electrostatic shock. Follow proper grounding
procedures.
The SITRANS LU SAM should be mounted in an area that is within the unit's ambient
temperature range and suitable for the specified enclosure. The front cover should be
accessible for viewing and have sufficient room to swing open; allowing for wiring and
setting switches.
The SITRANS LU SAM may be installed remotely from the main electronics to a maximum
loop length of 3,000 m (10,000 ft) when using 18 ga. wire.

Outline and Mounting

16 mm
(0.6")
mounting holes
(accessed under lid
4.3 mm (0.17") dia.
4 places)
Note:
Non-metallic enclosure does not provide grounding between connections.
Use grounded type bushings and jumpers.
Page 2
209 mm
(8.2")
172 mm
(6.8")
Suitable location for
conduit entrances.
Siemens Milltronics recommends using
a punch for making holes in enclosure
SITRANS LU SAM – INSTRUCTION MANUAL
lid screws
(6 places)
285 mm
(11.2")
267 mm
(10.5")
lid
enclosure
customer
mounting
screw
91 mm
(3.6")
7ML19985CF02