Measurement

Measurement
Note
Refer to the " Specifications " (page 12) to use this product in the sample condition. Otherwise it may
cause wrong measurement, and this product may break.
Measurement operation
1. Change valve to MEAS, referring to the " Selector valve operation " (page 7).
2. Perform measurement, referring to the instruction manual of the transmitter.
Note
The consumption of the KCl solution is 400 mL from 200 mL in one month. Replenish the KCl solution
every 2 weeks or 3 weeks.
Tip
When combining transmitter are HP-200 or HP-300,
It is possible to set a temperature coefficient. Refer to the instruction manual of the transmitter.
In the specification of the pH electrode, glass membrane error and reference electrode error are
not available.