8) Measure the flow out of the detector vent on the back panel and compare the measured value
with the target flow shown on the display. If necessary, use the ← key and the numeric keypad
to change the target pressure counts. Use the → key to update the new value. Wait for the
pressure counts to stabilize, at least 1 minute, and re-measure the flow at the detector vent.
Repeat until the measured flow is equal to the target flow ±2 cc/min.
9) When the measured flow equals the target flow, press "Enter" to save the span point.
10) If you make a mistake, hit Escape to exit out without saving the changes.