SLS Baseline 9000 FID
Total Gas Detector Manual
SLS Baseline 9000 FID
Document Number
Total Gas Detector Manual
GL-SLS-OEPS-L4-94
Originators
Lance Wheat
Rev 0
Mar 2016
Approval
No of Pages 38

Advertisement

Table of Contents
loading

Summary of Contents for Weatherford SLS Baseline 9000 FID

  • Page 1 SLS Baseline 9000 FID Total Gas Detector Manual SLS Baseline 9000 FID Document Number Total Gas Detector Manual GL-SLS-OEPS-L4-94 Originators Lance Wheat Rev 0 Mar 2016 Approval No of Pages 38...
  • Page 2 Note: The procedure in this manual should be performed in the order that the steps are written to properly operate the Baseline 9000 Total Gas Detector Page 1 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 3: Table Of Contents

    Sample ..........................18 Gas Calibration ..........................21 WellWizard Connection ........................27 Hardware Setup .......................... 27 Tensor Communication ....................... 28 Unhiding Baseline 9000 Channels ....................31 Bill of Materials ........................... 35 Page 2 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 4 Figure 35: Air Flow Value ..........................26 Figure 36: Baseline 9000 Static Units ......................27 Figure 37: Basline 9000 High Precision ....................... 27 Figure 38: Query Well ..........................28 Page 3 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 5 Figure 47: Unit Class ............................ 32 Figure 48: Hide Channel Status ........................32 Figure 49: Unhide Channel .......................... 33 Figure 50: Exit Channel ..........................33 Figure 51: Channel Category ........................34 Page 4 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 6: Configuration

    H2 volume generation capabilities before rig up to ensure the H2 generator being using has sufficient capacity to supply all equipment. 1.1 Baseline 9000/Agilent 490 1.1.1 Air Figure 1: Baseline 9000 with Agilent 490 Air Page 5 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 7: Sample

    1.1.2 Sample Figure 2: Baseline 9000 with Agilent 490 Sample Page 6 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 8: Fuel

    1.1.3 Fuel Figure 3: Baseline 9000 with Agilent 490 Fuel Page 7 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 9: Baseline 9000/Baseline 8900

    1.2 Baseline 9000/Baseline 8900 1.2.1 Air Figure 4: Baseline 9000 with Baseline 8900 Air Page 8 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 10: Sample

    1.2.2 Sample Figure 5: Baseline 9000 with Baseline 8900 Sample Page 9 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 11: Fuel

    1.2.3 Fuel Figure 6: Baseline 9000 with Baseline 8900 Fuel Page 10 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 12: Calibration

    Press Enter to select the displayed values or Escape to go back to the previous menu without saving the value. 4) Initial Flows should be set as: a. Fuel: 35 cc/min b. Air: 175 cc/min c. Sample: 8 cc/min Page 11 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 13: Flow Calibration

    4) Allow time to stabilize, minimum 2 minutes, and then press Enter to accept the zero point. NOTE: Keep the detector elevated to ensure no liquid enters from the detector vent. Page 12 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 14: Figure 10: Fuel Flow Span Point

    6) Reconnect the Fuel line to the TGD and apply 20 psi to the fuel inlet port. 7) Wait until the current pressure counts are equal to or close to the target pressure counts, minimum 2 minutes. Figure 11: Fuel Flow Press Cnts Page 13 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 15: Figure 12: Fuel Flow Value

    9) When the measured flow equals the target flow, press “Enter” to save the span point. 10) If you make a mistake, hit Escape to exit out without saving the changes. Page 14 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 16: Air

    5) The Span Point Calibration screen is now shown. Figure 15: Air Flow Span Point 6) Reconnect the Air line to the TGD and apply 20 psi to the air inlet port. Page 15 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 17: Figure 16: Air Flow Press Cnts

    1 minute, and re-measure the flow at the detector vent. Repeat until the measured flow is equal to the target flow ±2 cc/min. Figure 17: Air Flow Value Page 16 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 18: Figure 18: Air Flow Reading

    9) When the measured flow equals the target flow, press “Enter” to save the span point. 10) If you make a mistake, hit Escape to exit out without saving the changes. Page 17 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 19: Sample

    Keep the detector elevated to ensure no liquid enters from the detector vent. 5) The Span Point Calibration screen is now shown. Figure 20: Sample Flow Span Point 6) Reconnect the Sample line to the TGD.. Page 18 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 20: Figure 21: Sample Flow Press Cnts

    1 minute, and re-measure the flow at the detector vent. Repeat until the measured flow is equal to the target flow ±2 cc/min. Figure 22: Sample Flow Value Page 19 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 21: Figure 23: Sample Flow Reading

    9) When the measured flow equals the target flow, press “Enter” to save the span point. 10) If you make a mistake, hit Escape to exit out without saving the changes. Page 20 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 22: Gas Calibration

    2) Use the left arrow to delete the previous value if the machine does not already indicate 0.00 percent, and enter zero percent using the number pad. Press the green enter key to save. Figure 25: Zero Value Page 21 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 23: Figure 26: Span Selection

    4) Set the span value to 10.00% using the number pad on the right of the machine and the left arrow to delete previous numbers. Press the green enter key to save. Figure 27: Span Value Page 22 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 24: Figure 28: Calibrate Now

    If a zero calibration is not required, press the red key to skip this step and move onto the span calibration. Figure 29: Zero Calibration Page 23 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 25: Figure 30: Span Connection

    When the reading has settled, press the green enter key to save the calibration. This process may be skipped by pressing the red escape key if no span calibration is required. Figure 31: Span Calibration Page 24 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 26: Figure 32: Full Scale Test

    10) If the reading is high (i.e. 120%), the air flow must be adjusted. Go to setup on the main menu and scroll down to flows Figure 33: Flow Selection Page 25 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 27: Figure 34: Air Flow Selection

    100% calibration gas that is being applied. 14) Repeat the process until the machine reads 10% +/- 0.05 for 10% methane and 96% +/- 0.5% for 100% methane. Page 26 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 28: Wellwizard Connection

    2) Ensure that the Baseline 9000 is using high precision by navigating the menu on the machine and selecting the options as shown: Figure 37: Basline 9000 High Precision Page 27 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 29: Tensor Communication

    Figure 39: Enter Well Directory 4) Type “nano programs” to modify the programs file of the well Page 28 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 30: Figure 40: Programs File

    Figure 40: Programs File 6) Press ctrl+x to exit. 7) Press “y” to save 8) Press enter to return from the programs file 9) Enter “WWsetup” 10) Go to Well Management Page 29 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 31: Figure 41: Enable Well Logging

    11) Enable Well logging for the well. Figure 41: Enable Well Logging 12) Ensure that the device is transferring data by entering “wits_diag –p/dev/serX Figure 42: Wits Diag Page 30 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 32: Unhiding Baseline 9000 Channels

    Figure 44: Goto Channel 3) Enter “BLTGM”, this is mnemonic for the corrected baseline total gas value Figure 45: Mnemonic Entry 4) When prompted for a new channel name, press ENTER Page 31 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 33: Figure 46: Channel Name

    5) Unit Class will remain the same, press enter Figure 47: Unit Class 6) Using the arrow keys, scroll over to “Hide Channel” Figure 48: Hide Channel Status Page 32 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 34: Figure 49: Unhide Channel

    7) Press ENTER to change it from the default SET position Figure 49: Unhide Channel 8) Using the arrow keys, go to the EXIT and press ENTER Figure 50: Exit Channel Page 33 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 35: Figure 51: Channel Category

    9) Press Enter to confirm it is remaining under the same category Figure 51: Channel Category 10) Press ‘Shift+s’ to save the changes 11) Repeat for the following mnemonics: o BL Total Gas Raw (BLTG) Page 34 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...
  • Page 36: Bill Of Materials

    2231485 Tubing metallic 0.125+/-.005 OD 0.093 ID 0.016 THK 316 SS 12 feet • 2231484 Reducer tube 0.250 OD x 0.125 OD 316 SS 4 each • 2390212 Express Flow Panel 1 each Page 35 of 36 SLS Baseline 9000 FID Total Gas Detector Manual...