40.0° automatically.
When the temperature reaches to 40° and is able to hold
a constant temperature, the decimal digit begins to flicker.
After 20 minutes, the user can read the value on the
Thermometer and use the up and down keys to enter the
value of the thermometer.
Notes: Please read the actual value after 20 minutes' constant
temperature to ensure the calibration accuracy.
If the actual read out of thermometer is 39.6, modify the
temperature to 39.6 by pressing  or . Then press "Start/Stop"
to confirm the input value.
d) Then the instruments will the automatically begin to heat to 100℃. Once 100C is
reched and held as constact, the decimal begins to flicker. After 20 minutes, the
user can read the actual value from thermometer.
Notes: Please read the actual value after 20 minutes' constant temperature to ensure
the calibration accuracy.
If the actual readout is 101.5, modify the temperature display
to 101.5 by pressing  or . Then press "Start/Stop" to confirm
the input value.
Use the same method to calibrate the temperature deviation
in Block B.
Note:
Pressing "Set" and "" simultaneously during the temperature calibration
indicates exiting the temperature calibration program. The adjustments will be
cancelled and the changed value will be of no effect.