Electromagnetic Compliance
Manufacturer's declaration of electromagnetic immunity
The OML LINQ™ is intended for use in the electromagnetic
environment specified below. The customer or user of the OML
LINQ™ should assure that it is used in such an environment.
Immunity test
Electrostatic
discharge
(ESD) IEC
61000-4-2
Electrical fast
transient/burst
IEC 6100-4-4
Surge
IEC 6100-4-5
Voltage
dips, short
interruptions
and variations
IEC 61000-4-11
Power
frequency
(50/60Hz)
magnetic field
IEC 61000-4-8
IEC 60601
test level
± 8kV
contact
± 15kV air
No
connection
to mains
No
connection
to mains
No
connection
to mains
30 A/m
Com-
Electromagnetic
pliance
environment -
level
guidance
± 8kV
Floors
contact
wood, concrete or
ceramic tile. If floors
are
± 15kV
synthetic
air
the relative humidity
should be at least
30%.
ODFS Pace® (XL)
does not have a
mains power supply
30 A/m Power
magnetic
should be at levels
characteristic
typical location in a
typical
hospital or domestic
environment.
should
be
covered
with
material,
frequency
fields
of
a
commercial,
23

Advertisement