Service Result; Material Code; Coarse Sensitivity - Thermo Scientific Ramsey Oretronic IV User Manual

Tramp metal detector
Table of Contents

Advertisement

8
Clear Restart
Info
9
Software
Reset

Service Result

Base Address: H545
Access: Get Only
Persistent: No
This register contains an integer code value giving the result of the last service invocation. The encoding is
given in the table below.
Value
Meaning
0
Service Succeeded
1
Service Pending
2
Invalid Service Code
3
Other Error

Material Code

Base Address: H561
Access: Get/Set
Persistent: Yes
This is an integer code value which controls the delay from the end of the excitation pulse to the start of data
acquisition. It should be between 0 and 7. The time delay increases with code value.

Coarse Sensitivity

Base Address: H577
Access: Get/Set
Persistent: Yes
This parameter controls the receive A/D gain. The encoding is shown in the table below.
Value
ADC Gain
0
1
2
3
4
5
6
A count of restarts caused by watchdog reset and unexpected exceptions is maintained
in the EEPROM. This command resets this count back to zero.
Reset the hardware and software. When this command is sent the embedded
software stops updating the software watchdog. The reset is caused by a watchdog
timer expiration. This command can be used to test the software watchdog feature.
Data Type: UINT16
Length: N/A
Type: UINT16
Length: N/A
Min: 0
Max: 7
Type: UINT16
Length: N/A
Min: 0
Max: 6
1
2
4
8
16
32
64
Ramsey Oretronic IV Tramp Metal Detector REC 4479C