OML Microstim2 Instruction Manual page 26

Table of Contents

Advertisement

Technical
c.
Excessively high stimulation intensity - adjust appropriate level control.
The stimulator bleeps for 30 seconds then switches off:
a.
Low battery - replace battery if a non-rechargeable PP3 is used, recharge
battery if a rechargeable battery is used.
Specifications
Output amplitude:
Stimulation frequency:
Stimulation wave form:
Stimulation pulse width:
The Microstim2 is a Class IIa Medical Device within the definition of the European
Medical Devices Directive.
Expected Service Life and Shelf life
This is an indication of the life span of the parts (figures based upon daily use). Life
4
span will depend on usage.
Item
Microstim 2V2
Electrode socket
Electrode lead
Neurostimulation electrodes
PP3 Battery (9V) (non-rechargeable)
26
Maximum 100mA with 1KOhm output load.
Programmable (20 or 40Hz).
Nominally square with passive charge
balancing.
330µs maximum .
Expected service
life
6-12 Months
3-4 Weeks
6-12 Weeks
5 Years
2 Years
See expiry date
See expiry date
Shelf life
N/A
N/A
N/A