Important Information (Precautions)
Storage Temperature
Do not store outside of the temperature limits as detailed in the specification on
page 27 If the Microstim2 or accessories (excluding electrodes) have been stored
at the minimum (-20°C) or maximum (70°C) storage temperatures, allow at least 10
minutes in ambient temperature before use.
Use of Small Electrodes
The use of small electrodes (less than 32mm diameter) may result in higher charge
density especially when using high output settings (current, frequency) and may
lead to an increase in the likelihood of skin irritation. When possible use larger
electrodes. Always monitor skin condition closely.
Adverse Reactions
The patient should report any undesirable outcomes, malfunctioning of the device,
mistakes in using the device, or injury from the use of this device to the clinician
2
who provided it to them.
The clinician is responsible for reporting all adverse events to Odstock Medical
Limited or their local representative.
Residual Risk
Even though Odstock Medical Limited have taken all foreseeable steps, through
design and testing, to ensure this device is safe and reliable, there remains a small
risk of stimulation not being delivered when required.
Safe Disposal of Equipment
Care should be taken to dispose of parts and accessories of the device in the correct
manner. In order to protect the environment, the product and its accessories should
not be disposed of in household waste. Please dispose of as per local regulations.
Alternatively, equipment can be returned to Odstock Medical Limited for disposal