  • Page 1 MIRA Light Engine Instruction Manual Lumencor, Inc. Document Number 57-10001 www.lumencor.com...
  • Page 2 Lumencor utilizes regulatory model names for all certified and CE marked products. The regulatory model names are traceable to all regulatory documentation, third party reports and certifications. “Regulatory Model: Mira” is used as a representative model for all certified and CE marked Mira products. Emissions This equipment has been tested and found to comply with the limits of EMC directive 2014/30/EU and FCC part 15 (CISPR 11:+A1:2016).
  • Page 3: Table Of Contents

    Table of Contents 1. Introduction 2. Precautions and Warnings 3. Installation and Operating Instructions 4. Spectral Output 5. Product Specifications 6. Routine Maintenance and Trouble Shooting 7. Customer Support 8. Warranty MIRA Light Engine Instruction Manual...
  • Page 4: Introduction

    1. Introduction The Lumencor MIRA light engine is a compact solid-state light source for fluorescence microscopy and other bioanalytical applications. The MIRA light engine incorporates efficient thermal management, requiring no fan and providing quiet and vibration-free operation. This enables direct coupling to the fluorescence epi-illumination port of a microscope.
  • Page 5: Precautions And Warnings

    Mira light engine. It is imperative that the alternative power supply has output over-current protection, as the power input of the Mira is not fused. The DC power supply must have the AC power cord connected to a receptacle with a protective safety (earth) ground terminal.
  • Page 6 Attention: NE PAS placer de liquide sur la source lumineuse. Les liquides renversés peuvent endommager la source lumineuse. Caution: DO NOT drop the light engine. It contains glass optical components that could be damaged or misaligned by the shock produced by a drop onto a hard surface. MIRA Light Engine Instruction Manual...
  • Page 7: Installation And Operating Instructions

    DISCLAIMER: Lumencor shall not be liable for injury to the user or damage to the product resulting from the MIRA light engine being used in a way for which it was not intended and in complete disregard for any post- ed safety precautions and warnings.
  • Page 8 3.2 Installation NOTE: Any end-product/system incorporating or coupled to a Lumencor Light Engine shall be fully evaluated to verify all applicable safety and regulatory compliance requirements prior to use. Take necessary precautions to protect yourself and others from the high intensity light when turning on the unit.
  • Page 9 2. Move the COLOR control knob on the side panel of the MIRA) to the “ext” position [2]. 3. Press and hold the right button on the pod until a menu of light engines appears. Turn the rotary dial to select “MIRA”...
  • Page 10: Spectral Output

    Switching the gate connector to ground will then shut off light output. Restoring the input level to >3.3 V will restart light output. 4. Spectral Output The normalized spectrum shown below is representative of the MIRA light engine output. External bandpass filters can be used to restrict the wavelength range of each color channel. 5. Product Specifications The MIRA light engine must be operated and stored within the environmental conditions specified.
  • Page 11: Customer Support