High Flow Oxygen Therapy System Check
Single-limb Circuit
All checks run during a single-limb check are also run on a
dual-limb system check. There is no need to perform a system
check using a single-limb circuit if a dual-limb check was
performed and passed.
1. From Standby, select SYSTEM CHECK.
Note:
The Run System Check menu shows.
2. Select the single-limb circuit icon.
3. Attach the single-limb breathing circuit and all accessories
that will be used during O2 Therapy.
Note:
An inspiratory safety guard must be installed on the
inspiratory port and any accessories that will be used.
4. Select Start.
5. Select the information icon to see the System Check
Details menu.
a. As the system check is running the results of each check
will be displayed.
Note:
Wait for the System Check to finish. A green check mark
will display if the check passes. If the check fails, a red X is
shown, and a Help icon displays next to the failed check to
provide possible causes and help for troubleshooting. The
System Check results show the dual-limb circuit icon, the result
of the check, and the date and time of the check.
6. Select START O2 THERAPY to begin delivery of Oxygen.
7. Confirm Oxygen is flowing through the patient circuit
before connecting to the patient
(cont.)
1
5
5a
1
2
4
6
7