Drive does not operate
_____________________________________________________________________________
Bed sliding on incline
_____________________________________________________________________________
Bed continues rolling
after being stopped
_____________________________________________________________________________
Batteries need charged
Switch not in ON position
Short in directional control
switch
Bed not is lowest position
Tread wear on wheel
Dirty floor
Defective brake
Dirty floor
Bed is on irregular surface
Defective actuator motor
Ensure batteries are charged.
Switch
will
illuminate
power drive is on
Turn ON/OFF switch to OFF
then back to ON.
Lower bed to lowest position.
Replace wheel if less than 1/16
inch tread is left on wheel.
Clean floor.
Brake or motor will
need to be replaced.
.
Clean floor.
Due to the travel limits of the
motor suspension some
irregularities may result in the
wheel loosing traction with floor.