Do you have a question about the GALAXY 900HS and is the answer not in the manual?
Questions and answers
Steve Whitehead
April 18, 2025
On the 900HS, when using the pelvic drop, the left side doesn't go all the way back down (level with the other cushions/sections), causing the pelvic cushion to be crooked; and I have to use the adjustment handle to take it the rest of the way down. The right side does just fine. Also, if I use the adjustment handle on the left to release the pelvic drop, it releases both the left and right sides; but if I use the adjustment handle on the right, it only releases the right side, but not the left side. What might be going wrong and what do I need to look at to remedy this?
1 comments:
Mr. Anderson
May 9, 2025
The left pelvic drop on the Lloyd GALAXY 900HS may not lower completely due to incorrect tension adjustment. The Tension Control Knob on the left side increases tension when turned clockwise and decreases it when turned counterclockwise. If the tension is set too high, the drop may not release properly.
To resolve this issue:
1. Locate the left Tension Control Knob. 2. Turn the knob counterclockwise to decrease the tension. 3. Test the drop function to ensure it lowers completely. 4. Repeat adjustment if necessary. 5. Always check and test the drop tension before adjusting the patient.
Keep fingers away from the bottom of the cushion to avoid pinching.
CONTENTS OPERATIONS PAGE PARTS / ASSEMBLY PAGE Safety and Maintenance......1 - 2 Ankle Rest ..........Head Rest Cushion......... 3 - 5 Foot Stand ..........Chest Cushion......... Plastic Covers........Manual & Automatic Drops...... 7 - 9 Hydraulics..........28 - 29 Head &...
Page 4
WELCOME!! Lloyd Table Company, a proud supporter of today’s chiropractors, has established the finest line of chiropractic tables in the world. Table products unique in versatility and function, designed to meet your needs. SAFETY Concerned for a safe working and treatment environment, Lloyd Table Company asks that you follow the safety tips below.
Page 5
MAINTENANCE The following steps should be performed on a regular basis to maintain your Lloyd table in a smooth operating condition and to assure its long life and top quality appearance. An initial inspection and periodic future inspections should be made to insure that nuts and bolts have not loosened in shipping or during daily operations.
The following sections provide explanations on the operation of your Lloyd table. All references to right and left, will be in relation- ship to your view, as you face table from the foot end. All references to the front will refer to the head end of the table. Those to the rear will refer to the foot end.
Page 7
Standard - Deluxe Head To Elevate the Head Cushion The head cushion is assisted by a gas shock in raising the cushion. With a hand supporting the head cushion, turn the Elevation Knob (No. 2) on either side of the table upward to unlock.
Page 8
Optional - Deluxe Head with Drops #8 Tension Knob #7 Cocking Handle #9 Selector Knob (Right Side of Table) Operating the Straight / Toggle Drop: CAUTION: THE DROP CUSHION TENSION SHOULD BE CHECKED AND TESTED BEFORE ADJUSTING A PATIENT. NOTE - DROP SHOULD BE IN DOWN / DEACTIVATED POSITION BEFORE MOVING SELECTOR KNOB (NO.
Chest Cushion The chest cushion is designed to accommodate patients with large or small abdomens and has the versatility of being a stationary or floating cushion. #2 Spring Breakaway Knob #3 Spring Breakaway Tension Knob #1 Adjustable Chest Handle Standard - Adjustable Tilting Chest Height Adjustment of Chest Cushion Lift the rear corners of the cushion.
Page 10
Chest Cushion - Manual Drops #2 Tension Control Knob #2 Tension Control Knob #1 Cocking Handle Chest & Lumbar Drop - Manual Cocking the Drop Chest or Lumbar Cushion Pull up on the Cocking Handle (No. 1) to cock the desired drop cushion. Adjust the tension on either by turning the Tension Control Knob (No.
Page 11
Automatic Cocking Drops #1 Automatic Drop Selector Knob Cervical (Head Cushion) Drop - Automatic To Cock the Cervical (Head Cushion) Drop: Rotate the selector knob (No. 1) to “C” to select the Cervical (Head Cushion) drop. Step on the white strip to activate the auto cocking drop. CAUTION: THE DROP CUSHION TENSION SHOULD BE CHECKED AND TESTED BEFORE ADJUSTING PATIENT.
Automatic Cocking Drops To Cock the Pelvic Cushion Drop: Rotate the selector knob (No. 1) to “P” to select the Pelvic drop. Step on the white strip to activate the auto cocking drop. CAUTION: THE DROP CUSHION TENSION SHOULD BE CHECKED AND TESTED BEFORE ADJUSTING PATIENT.
Ankle Rest #2 Adjustment Lever #1 Lock Knob Ankle Rest Adjustment To Adjust Height of Ankle Rest Loosen the Lock Knob (No. 1) Simply lift on the sides of the Ankle Rest Cushion. Once the desired height is obtained the cushion will automatically lock into position when released.
Tilt Safeguard Left Arm Rest Tilt Safety Switch Tilt Safety Switch The tilt safety switch is designed to prevent unauthorized operators from lowering the table from a Vertical position. The table cannot be lowered until the tilt Safety Switch is activated. The switch is located (shown above) on the left side, behind the arm rest cushion.
Foot Pedal Control Directional Pedal The directional pedals have the dual function of raising and lowering table in elevation and hylo. The directional pedals can be operated in manual and auto mode as described on page 13 under auto / manual. Hylo Corded Side - Lowers the table from Vertical to horizontal.
Height Select Panel #1 Height Select #2 Auto / Manual #3 Horz. Down / Tilt Up Switch #4 Horz. up / #5 Beeper / Mute Tilt Down Switch Switch Height Select (No. 1) The height select panel consists of nine height select buttons whereby the operator can preselect a desired height. The buttons range from 19 to 36 inches, at two inch increments.
BASE FRAME ASSEMBLY ITEM PART N0. DESCRIPTION QTY. 55110 FRAME, BASE, 900 58026 WHEEL PLATE ASSEMBLY (RH), 19” 900 58027 WHEEL PLATE ASSEMBLY (LH), 19” 900 21910 WASHER, FLAT, 1/2 ID X 1 OD X 0.095 20542 BOLT, HEX HD, 1/2 - 20 NF X 2 1/4 (CHROMED) 20115 BOLT, HEX HD, 1/4 - 28 NF X 3/4 21020...
Page 20
SCISSORS YOKE ASSEMBLY GALAXY 900 HS ITEM PART N0. DESCRIPTION QTY. 55210 YOKE FRAME ASS’Y 22900 BUSHING, BRASS, 7/8ID X 1 ¼ OD 30610 SLEEVE, STEEL, ½ ID X 7/8 OD X 1 ¼ 21910 WASHER, FLAT, ½ ID X 1 OD 20510 BOLT, HEX HD, ½-20 UNF X 1 ¼...
Page 21
SCISSORS FRAME ASSEMBLY GALAXY 900HS ITEM PART N0. DESCRIPTION QTY. 55311 H - FRAME ASSEMBLY, DOUBLE 55330 H - FRAME ASSEMBLY, SINGLE 50341 BAR, LINKAGE, 19” 58113 TONGUE ASSEMBLY, 19” 50356 BOLT, HEX HEAD, 1/2 - 20 NF X 2 (SHAVED HEAD)
DELUXE HEAD FRAME ASS’Y ITEM PART N0. DESCRIPTION QTY. 56051 FASTENER PLATE ASS’Y 57221 ADJUSTMENT ARM 55929 CENTER BAR 21616 PLASTIC FLAT WASHER, ¼ ID X ¾ OD 24020 HANDLE 22050 RD. HD. MACHINE SCREW, ¼-20 UNC X 2 ½ 21450 HEX.
Page 23
DELUXE HEAD CUSHION ASS’Y NOTE: ITEM #20 IS USED WITH THE STRAIGHT DROP HEAD CUSHION ITEM PART NO. DESCRIPTION 40110 CUSHION, HEAD - LEFT HAND 40130 CUSHION, HEAD - RIGHT HAND 29999 HEADREST PAPER 30400 SLIDE BLOCK, FRONT 22009 OVAL HD. MACHINE SCREW, #10-24 NC X 1 1/2 57174 SPACER, 3/16 ID X ½...
Page 24
I - FRAME ASSEMBLY DELUXE, FORWARD MOTION & FLEXION HEAD Page 20...
Page 25
I - FRAME ASSEMBLY DELUXE, FORWARD MOTION & FLEXION HEAD ITEM PART N0. DESCRIPTION QTY. 55553 I-FRAME ASSEMBLY 50872 GUIDE ROD, 1/2 RD X 7 7/8 21620 WASHER, FLAT, 1/4 ID X 9/16 OD 22013 SCREW, MACHINE, BUTTON SOC HD, 1/4 - 20 NC X 1/2 21975 SCREW, MACHINE, PAN HD, #6 - 32 NC X 1/2 21010...
Page 26
STANDARD CHEST CUSHION ASS’Y SPRING BREAKAWAY Page 22...
Page 27
DELUXE CHEST CUSHION SPRING BREAKAWAY ITEM PART N0. DESCRIPTION QTY. 40210 CUSHION, CHEST - STANDARD 40230 CUSHION, CHEST - WIDE 51010 BRACE, CUSHION (LH) 51030 BRACE, CUSHION (RH) 20022 BOLT, HEX HEAD, 1/4 - 20 NC X 1 20361 BOLT, HEX HEAD, 3/8 - 16 NC X 1 3/4 21850 WASHER, FLAT, PLASTIC, 3/8 ID X 1 OD 55945...
Page 28
PELVIC ASSEMBLY GALAXY 900 HS ITEM PART N0. DESCRIPTION QTY. 40310 CUSHION, PELVIC, NARROW 40329 CUSHION, PELVIC, NOTCH, WIDE 40330 CUSHION, PELVIC, WIDE 51310 ANGLE, FRONT PELVIC SUPPORT, 1/4 x 1 x 1 x 8 1/16 51330 SLIDE ROD, 23/32 x 19 3/4 56350 FRAME ASSEMBLY, PELVIC 22090...
ANKLE REST GALAXY 900 HS ITEM PART N0. DESCRIPTION QTY. 40430 ANKLE REST CUSHION, STANDARD 40410 ANKLE REST CUSHION, WIDE 56537 EYE-ROD, 7/16 DIA X 7 11/16 56516 SUPPORT ROD ASSEMBLY 22468 SCREW, SHEET METAL, #10 X 7/8 22922 COLLAR, SET, 5/16 ID X 5/8 OD 22090 SET SCREW, #10 - 32 NF X 3/16 20199...
HYDRAULIC GALAXY 900 HS ITEM PART N0. DESCRIPTION QTY. 623500 PUMP & MOTOR 45391 HOSE, HYD, 3/16 X 80 1/2 (MOTOR TO TILT CYLINDER) 45396 HOSE, HYD, 3/16 X 91 1/2 (MOTOR TO EXT / RET CYLINDER 45300 HOSE, HYD, 3/16 X 30 1/2 (MOTOR TO ELEVATION CYLINDER) 21810 WASHER, FLAT, 3/8 ID X 11/16 OD 20320...
STRAIGHT DROP HEAD ASS’Y Items 49, 50, 51, 52, 53 and 54 are required only for Auto Cocking Cervical Tables. Page 32...
Page 37
STRAIGHT DROP HEAD ASSEMBLY ITEM PART N0. DESCRIPTION 56011 FASTENER PLATE ASS’Y, ’89 STRAIGHT DROP HD 57201 OUTSIDE DROP FRAME ASS’Y, ’89 FM HEAD 55850 TEE WEDGE ASSEMBLY 22017 SCREW, MACHINE, FLAT HEAD, ¼ - 20 UNC X ½ 55929 CENTER ROD ASSEMBLY 21616 WASHER, PLASTIC, ¼...
Page 38
NOTE: ITEMS 28, 33, 34, 35, 36, 37, 38 & 40 ARE USED ONLY FOR TABLES WITH AUTO COCKING CERVICAL. Page 34...
Page 39
FORWARD MOTION HEAD FRAME ITEM PART NO. DESCRIPTION 56059 OUTSIDE DROP FRAME ASS’Y 57201 OUTSIDE HEAD FRAME ASS’Y 55850 TEE WEDGE 57221 ADJUSTMENT ARM, HEAD CUSHION 57175 LINKAGE BAR, 1/8 X 1/2 X 3 9/16 21995 PAN HD. MACHINE SCREW, #10-24 NC X 1/2 22860 BUSHING, 3/16 ID X 1/4 OD 30722...
Page 45
SLIDE FRAME ASSEMBLY FLEXION HEAD ITEM PART N0. DESCRIPTION QTY. 55778 SLIDE BRACKET ASS’Y 55794 LOCK FRAME ASS’Y 22865 BUSHING, ¼ ID X 5/16 OD 22657 ROLL PIN 1/8 X 3/8 22660 LOCK KNOB 22031 SOCKET HD SCREW, ¼-20 UNC X 1 ¼ 30750 COMPRESSION SPRING 55762...
Page 46
ROTATION BLOCK ASSEMBLY FLEXION HEAD Item 56 - Flexion Handle Kit #634010 Part No. Description 55838 Handle Tube 55836 Side Ass’y (RH) 55837 Side Ass’y (LH) 24023 Handle Grip 24024 Adjustable Thumb Handle 20188 Socket Hd. Bolt #10-24 22410 Pan Hd. Screw 24060 Buckle, Plastic 28108...
Page 47
ROTATION BLOCK ASSEMBLY FLEXION HEAD ITEM PART N0. DESCRIPTION QTY. 55777 ROTATION FRAME ASSEMBLY, FLEXION HEAD 55773 PIVOT ARM ASSEMBLY, FLEXION HEAD 55963 STABILIZER BRACKET ASS’Y, SWIVEL HEAD 55788 GUIDE ROD, CRADLE ASS’Y, FLEX HD, 3/8 RD X 7 5/8 21610 WASHER, FLAT PLASTIC, 0.26 ID X ½...
DROP CHEST & LUMBAR CUSHION Note: Item #40 is used only on table’s with Breakaway Chest. Note: For Detail Parts on Breakaway Chest refer to Deluxe Chest Cushion (page 22 - 23). Page 44...
Page 49
DROP CHEST and LUMBAR ASSEMBLY For use with STANDARD or BREAKAWAY Type Table ITEM PART N0. DESCRIPTION QTY. 57309 CHEST & LUMBAR FRAME ASSEMBLY 40510 CUSHION, DROP CHEST 40530 CUSHION, DROP LUMBAR 57330 MOUNTING PLATE ASS'Y 57416 COCKING HANDLE BAR (RH) 57436 COCKING HANDLE BAR (LH) 30420...
DROP CHEST ASSEMBLY FULL CUSHION ITEM PART NO. DESCRIPTION QTY. 40235 CHEST CUSHION ASSEMBLY 57428 TEE WEDGE ASSEMBLY, FRONT, NEW DROP 57429 TEE WEDGE ASSEMBLY, REAR, NEW DROP 20022 BOLT, HEX HEAD, 1/4 - 20 NC X 1.00 20332 BOLT, HEX HEAD, 3/8 - 16 NC X 1.00 57435 BAR, STABILIZER, DROP CHEST, 1/4 X 5/8 X 6 13/16 21210...
Page 57
T-WEDGE DROP PELVIC ASS’Y Z-FRAME ITEM PART N0. DESCRIPTION QTY. 40308 PELVIC CUSHION 56158 TEE WEDGE ASS’Y 20023 HEX HD BOLT, ¼-20UNC X 1 ¼ 20332 HEX HD BOLT, 3/8-16UNC X 1.00 20360 HEX HD BOLT, 3/8-16UNC X 1 ½ 22812 BALL BEARING, 3/8ID 21210...
Page 61
HYDRAULICS - AUTO DROPS CERVICAL, DORSAL, LUMBAR & PELVIC ITEM PART N0. DESCRIPTION QTY. 623580 PUMP & MOTOR, AUTO DROP 45391 HOSE, HYDRAULIC, 3/16 X 78 1/2 (MOTOR To JOINT) 23047 COUPLER, MALE, 1/4 JIC To 1/4 JIC 45294 HOSE, HYDRAULIC, 3/16 X 24 1/2 (JOINT To SELECTOR VALVE) 31010 VALVE, SELECTOR 45212...
Page 62
GALAXY 900 HS WIRING Junction Logic Board 4 - Pin Connector # 45032 Electrical Unit # 45017 6 - Pin Connector White - Motor Connector Yellow - AC Dist Box Quik White - 24v tran #4 Connect Motor Connect Yellow White - 5v Tran #3 5 Volt Blue...
Page 63
GALAXY 900 HS WIRING Orange - 24v Tran #1 Yellow - 5v, Tran #4 Term inal Strip (# 38360) White, See Legend Below Motor Relay 24v Tran #4 Outside Inside Green Black Black Wire AC Dist Box Quik Connect Red White Capacitor Green Wire...
Page 64
GALAXY 900 LIMIT SWITCHES 900 HS Tilt Limit Switch 900 HS Lower Tilt Limit Switch Wire (Com) Black Wire (NO) Black Wired to COM & NO 900 HS Activator Footplate Switch Black To Tilt Limit Sw itch Activator Wire Black Table Foot End Table Head End N.O.
Page 69
For your convenience a table maintenance schedule has been provided to keep track of all repairs and parts purchases. Date Part Number ***Always include the table’s serial number when calling Lloyd Table Company Service Department*** Telephone: 319-455-2110 Toll Free: 800-553-7297 Fax: 319-455-2166 Email: service@lloydtable.com...
Need help?
Do you have a question about the GALAXY 900HS and is the answer not in the manual?
Questions and answers
On the 900HS, when using the pelvic drop, the left side doesn't go all the way back down (level with the other cushions/sections), causing the pelvic cushion to be crooked; and I have to use the adjustment handle to take it the rest of the way down. The right side does just fine. Also, if I use the adjustment handle on the left to release the pelvic drop, it releases both the left and right sides; but if I use the adjustment handle on the right, it only releases the right side, but not the left side. What might be going wrong and what do I need to look at to remedy this?
The left pelvic drop on the Lloyd GALAXY 900HS may not lower completely due to incorrect tension adjustment. The Tension Control Knob on the left side increases tension when turned clockwise and decreases it when turned counterclockwise. If the tension is set too high, the drop may not release properly.
To resolve this issue:
1. Locate the left Tension Control Knob.
2. Turn the knob counterclockwise to decrease the tension.
3. Test the drop function to ensure it lowers completely.
4. Repeat adjustment if necessary.
5. Always check and test the drop tension before adjusting the patient.
Keep fingers away from the bottom of the cushion to avoid pinching.
This answer is automatically generated