Appendix A. Fault Finding
Error Message
Probable cause
!021 Bad
Error occurred in memory check sum.
Checksum
!022 Bad
Erroneous character detected in the
Message
middle of a message.
!023 Bad Cal
Divide by zero error occurred in
Pres
calibration calculation.
Copyright 2010 Baker Hughes Company.
30 | TERPS Instruction Manual–English
Table A-1: Fault Finding
Action
Power cycle sensor, if fault remains return
to Druck for repair or replacement.
Check and re-send message.
Check data and repeat calibration.