H - Change Slope - Baker Hughes Druck 8000 Series Instruction Manual

Trench etched resonant pressure sensor
Table of Contents

Advertisement

Reply:
Reply:
Command:
Non-interactive calibration requires the C command to be used twice, i.e. two calibration
pressures must be entered for a straight line fit. Proceed as follows:
Command:
where:
<PIN>
Apply pressure 1, when stable send:
Command:
Note: Sending *C,PIN,1,?<CR> causes the current stored data for pressure 1 to be sent.
Apply pressure 2, when stable send:
Command:
Query:
Reply:
Query:
Reply:

6.7.2 H - Change Slope

This command changes the slope of the output function of the transducer.
INFORMATION
Do not use the H and S commands together.
Use one command and then assess the result before continuing.
"Pressure offset = <new pressure offset>"<CR>
"Accept these values? (Y/N) =>(Enter Y or N)
(Y answered) "EEPROM updated"<CR>
(N answered) "Current settings left unchanged"<CR>
(If invalid data entered) "!023 Bad Cal Pres"<CR>
<SPACE>*C,<PIN>,mode,?<CR>
<SPACE>*C,<PIN>,1,?<CR>
is the PIN of the transducer and '1' is the first pressure value.
<SPACE>C,<PIN>,1,<CR>
<SPACE>C,<PIN>,2,<CR>
<SPACE>C,<PIN>,mode?<CR>
<Measured>,<Applied><Got first point (Yes/No)><CR>
<SPACE>*C,?<CR>
"Measured Pressure = <Measured>"<CR>
"Applied Pressure = <Applied>"<CR>
"First Point Done = <Yes/No><CR>
Proceed with caution when using this command.
Copyright 2010 Baker Hughes Company.
English–TERPS Instruction Manual | 21