Lift Motor Replacement (PR-906 only)
1.
Unplug the transformer from the wall outlet.
2. Gently push the chair over onto the footrest.
3. Unplug the motor from the hand control extension and the transformer.
4. Remove the four (4) bolts; two (2) per side from the front motor mount and retain hardware for
reassembly.
5. Remove the hitch pin and the clevis pin from the rear motor mount that fastens the bottom of the
motor to the lift frame. Retain hitch pin and clevis pin for reassembly.
6. Remove the motor from the lift frame.
7. Align the new motor into place and use the hitch pin and clevis pin to fasten the bottom of the new
motor to the lift frame.
8. Use the four (4) bolts and hardware retained in step 4 to fasten the top of the lift motor to the lift
frame.
9. Connect the new motor to the hand control extension and the transformer.
10. Turn the chair over onto the lift frame.
11. Plug the transformer into the wall outlet and test.
41
Front Motor Mount
(Side view of bolts)
906 LIFT MOTOR
Lift Frame
(Bottom section)
REVA_082012
Rear Motor Mount
(Clevis & Hitch pins)