3.2

Mechanical Safety

SI
Toothed belt of the W-drive
SI
Toothed belt of the O-drive
Visually inspect the toothed belts of the drive for wear.
Cracks on the back or at the tooth base; replace toothed belts if necessary.
Ensure that the clamping pieces of the toothed belts are secure.
SI
X-ray tube assembly
SI
Collimator
SI
Image intensifier with camera head
Ensure that the screw connections of the X-ray tube assembly on the C-arm are secure.
Ensure that the collimator is securely attached to the X-ray tube assembly.
To check the mounting of the collimator, remove the large rectangular cover of the radi-
ation protection housing and then reinstall it. The collimator may not rotate at the hous-
ing flange.
To check the image intensifier mounting, remove its rubber cover and the protective
cover of the camera.
Check the mounting of the image intensifier.
Check the mounting of the camera at the image intensifier.
Reinstall covers.
SI
Unit mounting
SI
Safety distance of the unit
Use a torque wrench to check the attachment of the table to the base plate. Ensure that
8 x Allen screws (at the lifting columns) M8 x 40 with 24 Nm ± 10% 4 x external hexa-
gons M10x25 are secure.
Check the maneuverability ranges of the unit, the O-W -X-Y-Z-drive during lithotripsy
and urology operation. Maintain the safety distances in the end positions of the unit
movements.
SI
Urology insert
Insert the urology insert and ensure that it is secure.
Ensure that S22 is properly switched.
Adjust switch S22 if necessary.
If needed, adjust both guide pieces of the urology insert. For this purpose, insert the
urology insert and press both guide pieces toward the outside.
Tighten the screws.
SI
Shock wave heads
Check the maneuverability of the shock wave system support (W-drive) from the
parked position to the working position and back.
Check the limited range by moving unit G1 in the table cutout.
SI
Unit grip strip
Ensure that the unit grip strip is secure.
LITHOSTAR Multiline
RXL2-120.101.01.06.02
08.05
Safety inspections
Page 16 of 52