46
10.3
Isocenter
PMA
Adjustment
•
Check the isocenter at the table with the adjustment phantom and readjust it in the case
of deviations outside of the tolerance.
•
Check the adjustment of the shock wave head.
(See "Adjusting the isocenter" in the adjustment instructions).
•
If the ultrasound option is available, also check the isocenter with ultrasound. (See in-
structions RXL-120.091....).
LITHOSTAR Multiline
RXL2-120.101.01.06.02
08.05
System adjustment
Page 46 of 52
CS PS 24
Siemens AG
Medical Solutions