Ⅵ. SYSTEM SPECIFICATIONS

Concentrator Specifications

Dimensions
Weight
User Interface
Sound Level
Warm-Up time
Oxygen
Concentration
Flow Control
Settings and
Pulse Volumes
Breathing
Frequency
Inspiratory
Trigger
Sensitivity
Maximum Outlet
Pressure
Use Mode
System Specifications
L/W/H: 8.70in (22.1cm) / 3.35in (8.5cm ) / 6.30in (16.0cm)
4.37 pounds / 1.98Kg (with battery installed)
2.8 inch LCD color display screen
49 dB(A) (on setting 2)
2 minutes
90% (-3% +6%) at all settings
Settings
1
Breath
Rate
10
21
15
14
20
10.5
25
8.4
16.8
30
7
35
6
40
5.3
10.5
±15% at STPD*
+/-25% over the rated environmental range
*STPD is 101.3 kPa at an operating temperature of 20°C,
dry
10 to 40 BPM
≤ 0.12 cm H
O
2
25 PSI
Continuous Use
30
2
3
4
Pulse Volumes (ml)
42
63
84
28
42
56
21
31.5
42
25.2
33.6
14
21
28
12
18
24
15.8
21
5
100
66.7
50
40
33.3
28.6
25