Ⅴ. MAINTENANCE AND CLEANING OF P2

Cleaning the Case

The outside case should be cleaned using a cloth dampened with a solution of mild
detergent and water.
Do not allow liquids into any of the controls, the interior of
CAUTION
the case, or the oxygen tubing connector. If this occurs,
contact your equipment provider for assistance.
Do not use alcohol, isopropyl alcohol, ethylene chloride or
WARNING
petroleum-based cleaners on the case or particle filters.

Cannula Replacement

The nasal cannula is designed for disposable use. Replacements can be purchased
from your doctor and/or concentrator equipment providers.
Nasal cannulas should be rated for 5 liters per minute to
CAUTION
ensure proper patient usage and oxygen delivery.

Filter Cleaning and Replacement

Filters are designed to provide adequate air flow through the device and are located at
the front of the P2.
Pre-Filter - part # FI-P200
This pre-filter must be cleaned on a weekly basis to ensure
adequate air flow through the device. Clean the pre-filters with a
mild liquid detergent and water. Be sure that the filter is completely
dry before reinstalling.
It may be necessary to clean the pre- filters more
CAUTION
frequently in dusty environments or conditions.
Maintenance and Cleaning of P2
26