Description of the P2 Oxygen Concentrator

Alerts Table

Alarm item
Battery cycles > 500
Battery
or battery health <
Exhausted
50%
Replace Sieve
Sieve bed expired or
Bed
Sieve bed chip error
Low Input
Adapter input
Voltage
< 17.0v
No breath detected
Absence of
continuously for
Breath
more than 15
seconds
Concentration < 87%
Oxygen
continuously for
concentration
more than 300
<87%
seconds
5%≦RSOC≦20%
Low Battery
without adapter
Concentration < 50%
Oxygen
continuously for
concentration
more than 300
<50%
seconds
Breath Sensor
Breath Sensor failed
Fail
Alarm condition
Alarm only
Alarm only
Auto-switch to
battery until the
adapter input >
18v
Alarm only
Alarm only
Alarm only
Auto-shut down
after 30 seconds
Auto-shut down
after 30 seconds
16
System process
Screen Display
Battery is
exhausted.
Replace battery
Replace Sieve
Bed
Please contact
provider
Low Input Voltage
Please check
adapter
Absence of Breath
Please check
cannula
Low O
Please contact
provider
Low Battery
Please charge
Low O
Please contact
provider
Breath Sensor
failed
Please contact
provider
:< 87%
2
:< 50%
2