3.3 Preset mode

Preselection of a reference value (preset)
The two reference values (REF I and REF II) can be preselected differently and up to a maximum
preset value of ± 2999.99 mm / ± 89.9995 IN.
1. Select active reference value (in reference mode).
2. Select preset mode (press and hold [mode] until the status display shows 'PRESET').
Entering the value you want to change
Press [mode] briefly one or more times to move the cursor under the number you want to change.
After the last number the cursor moves under the +/- sign again.
Changing a number
Press [set] briefly one or more times to increase the underlined number by one unit. Press and hold
the [set] button to move through the numbers automatically until the required value is reached.
Changing the preset +/- sign
Move the cursor under the sign and change the sign by pressing [set] briefly.
Confirming the preselected reference value (preset)
Press and hold the [mode] button until the 'PRESET' status display disappears.
Recommendations for use
For REF I, preselect the reference value 0.000 as a constant preset value, and use REF II for the
preselection of different reference values.

3.4 Connecting the serial interface:

To connect the serial cable (option) the cover has to be removed. The label of the connector must look
downwards to plug it in.
1

3.5 Special note:

-
Contaminations of the device could affect the measuring process
-
To clean a contaminated device please use a dry towel. If it is not sufficent please use a towel
with a neutral solvent. To prevent any damages don't use volatile solvent, as it causes
damages on plastic parts.
-
At regular intervals please lubricate the plug gauges with a resin free oil. It will prevent the
gauge from corrosion.
-
Opening the device will leed to loose the warranty claim.