  • Page 1 Installation and Operation Instructions G-Series Patient Lift (400 lbs & 700 lbs)
  • Page 2: Table Of Contents

    Technical Specifications for Vancare G-Series Lift Anatomy of the G-Series Lift Basics in Transferring a Patient Installing the G-Series Lift in the Vancare PLP Track 9-10 Installing the Endstop on the Vancare Track Connecting the Carry Bar to the Lift Strap...
  • Page 3: Introduction

    700 lbs (318 kg) The Vancare G-Series Lift is a fixed lift and can be installed in a variety of track profiles (the trolley can be customized to fit these track profiles) and it also works seamlessly with the Vancare G-Series Pendant (PLP) Sytem.
  • Page 4: Symbols Used In This Manual

    CAN/CSA-C22.2 No. 60601-1:08 Medical Electrical Equipment – Part 1: General requirements for Standard(s): basic safety and essential performance. ISO10535:2006 Hoists for the transfer of disables persons – Requirements and Test Methods Vancare G-Series Lift Product: G-Series Lift Brand Name: Models: 400 lbs and 700 lbs 3.24”...
  • Page 5: Safety Instructions

    • Contents of this manual are subject to change without prior notice to users. Keep for future reference. • Never place the Vancare G-Series Lift, track/PLP and sling(s) in control of a person who has not been properly trained in the use and care of this equipment.
  • Page 6: Technical Specifications For Vancare G-Series Lift

    Tec nical S ecifications for Vancare G-Series Lift Three G-Series Lift Weight Capacities Single Lift Lift Up to G-SERIES 400 400 lbs / 181 kg Lift Up to Single Lift G-SERIES-700 700 lbs / 318 kg Lift Tandem Lifts Lift Up to...
  • Page 7: Anatomy Of The G-Series Lift

    Lifts up to 700 lbs! just weighs 10 lbs Durable all metal gears Bright LED indicators Constructed of solid steel Quick magnetic hand control connection Emergency lowering/ stop feature Capacitive touch Rubber emergency strap hand control for easy cleaning www.vancare.com...
  • Page 8: Basics In Transferring A Patient

    Working with the Vancare G-Series Lift: The Vancare G-Series Lift moves freely in the track system and does not have any special requirements for space or power in connection with moving. Attention can this be fully focused on the user’s functional level and the caregivers technique.
  • Page 9: Installing The G-Series Lift In The Vancare Plp Track

    • Only persons who have received competent instruction regarding the use of the lifting equipment and fitting of slings should use the Vancare G-Series Lift. Important: Plan the move and avoid leaving the patient in the sling unattended. Before lifting, check that the patient is completely free of his/her surroundings.
  • Page 10 6. Fasten all end stops tightly using an adjustable wrench. NOTE: If you are installing the Vancare G-Series Lift in an existing track system you must ensure that the max load of the track system is equal or higher than the max load of the Vancare G-Series Lift.
  • Page 11: Installing The Endstop On The Vancare Track

    En sto on t e Vancare Trac 1. Use a 3/16 Allen key to secure the endstop to the track. VancareTrack Endstop Connecting the Carry Bar to the Lift Strap 1. Hold the carry bar and press the button using the thumb as shown.
  • Page 12: Plate Operating The G-Series Lift

    Operating the G-Series Lift The Vancare G-Series Lift is switched on automatically when a button on the hand control is pressed. The Vancare G-Series Lift switches off automatically after approximately three minutes without activation. "UP" button "DOWN" button Charging the G-Series Lift The charger contacts with two metal charging strips located inside the track.
  • Page 13: Led And Buzzer Functions

    After an emergency, the lift must be inspected prior to restoring to use. In order to restore power to the lift, the tab must be pressed back into the lift. Do not pull the red cord forcefully. For assistance after an emergency contact: Vancare Service info@vancare.com or 1.800.694.4525 www.vancare.com...
  • Page 14: Emergency Lowering

    NOTE: The emergency lowering button does not provide a raising function. The failure of any of the lowering devices should be reported to Vancare. In an event of emergency when normal lowering system of the lift malfunctions and the “emergency lowering”...
  • Page 15: Cleaning And Disinfection

    Take great care to ensure that no liquids get inside the Vancare G-Series Lift. The lift is not drip proof or water tight. Failure to protect the lift from liquids may result in damage to the lift and may cause personal injury.
  • Page 16: Inspection And Maintenance

    Inspection and Maintenance Prior to using the Vancare G-Series Lift, the inspections should be conducted per the following schedule: Item Before Use Every Month Annual Every Two Years G-Series Lift • Ensure that end stops are installed • Inspect strap for wear or fraying •...
  • Page 17: G-Series Track Systems And Support Structure

    White powder coated extruded aluminum Triple TRK-TRP-240 240 x 2.63 x 5.75" / 6100 x 67 x 146 mm 12.4 lbs / 5.65 kg White powder coated extruded aluminum Custom Custom lengths in any track type available upon request Vancare...
  • Page 18: Environmental Conditions

    Relative atmospheric humidity Air pressure 700 hPa 1060 hPa References on the Package +50 C° 20% - 90% 700hPa - 1060hPa -10 C° Temperature range Atmospheric Air pressure for transport humidity for transport and storage for transport and storage and storage www.vancare.com...
  • Page 19: Electro-Magnetic Compliance Data For Vancare G-Series

    Electro-Magnetic Compliance Data for Vancare G-Series Lift Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment speci ed below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
  • Page 20 Electro a netic om liance ata For Vancare G-Series Lift Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment speci ed below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
  • Page 21 Electro a netic om liance ata For Vancare G-Series Lift Guidance and Manufacturer’s Declaration - Electromagnetic Emissions The [EQUIPMENT or SYSTEM] is intended for use in the electromagnetic environment speci ed below. The customer or the user of the [EQUIPMENT or SYSTEM] should assure that it is used in such an environment.
  • Page 22 Disposal • The Vancare G-Series Lift doesn’t contain any dangerous goods. • The components of the Vancare G-Series Lift should be properly disposed at the end of its shelf-life. • Make sure that the materials are carefully separated. • The electrical conducting boards should be submitted to an appropriate recycling proceeding.
  • Page 23: Lift Warranty

    The primary purpose of our ceiling lift system is to safely lift, transfer and reposition a patient with a little effort as possible for the caregiver, regardless of the room type. Vancare s G-Series Lift is easy and safe to use for caregivers as well as patients.
  • Page 24 Vancare | www.vancare.com 1515 1st Street Aurora, NE, 68818 Toll Free Tel: 1.800.694.4525 Tel: 402.694.4525 Fax: 402.694.3994 Email: info@vancare.com...

This manual is also suitable for:

Golift 400Golift 700Golift 1000