After Surgery
ForceTriad User's Guide
Reactivate
A four-pulsed tone will sound when the Reactivate screen is displayed.
If this message appears the user should:
1.
Release the footswitch pedal or hand switching button.
2.
Reactivate the seal cycle without repositioning the instrument.
Endpoint not reached – Additional time and energy are needed to complete the
fusion cycle.
Seal cycle interrupted – The seal cycle was interrupted before completion. The
handswitch or footswitch was released before the end tone activated.
Reactivate / Energize – Reactivate the seal cycle without removing or
repositioning the instrument.
Disconnect the instruments
Turn off the energy platform.
1.
Disconnect all instruments from the front panel.
2.
•
If the instrument is single use only (disposable), dispose of it according to
the procedures for your institution.
•
If the instrument is reusable, clean and sterilize it according to the
manufacturer's instructions for use.
Disconnect and store any footswitch(es) used.
3.
After Surgery
6-9