Fast Calibration (Orientation) - Rangevision NEO Setup And Calibration

Table of Contents

Advertisement

After all the necessary snapshots are taken, wait for the result of calibration procedure.
From now on the
resulting accuracy of calibration will be specified on the home screen
in pixels. A good result is accuracy
result, repeat the calibration, carefully checking the plates position and brightness. If is is
necessary, repeat calibration without blue light and filters.
Some advice on calibration:
do not change the distance from the center of the plate to the scanner when
turning the plate. The exception are only snapshots in positions 10 and 11,
do not rotate the plate at a very high angle. Ensure that all marks on the plate are
visible from both cameras, when you turn the table,
be careful with calibration plates! It is not permitted to contaminate or perform
any mechanical damage to the surface with the marks. After the use store the
plates in designated case,
before the calibration procedure ensure that cameras and cables are secure in
the designated position.
Fast calibration (orientation)
no worse (not higher) than 0.2
pix. If you get a higher
13