ILX Lightwave OMH-6722B User Manual page 20

Silicon power / wavehead
Table of Contents

Advertisement

C H A P T E R
4
OMH-6722B
Downloaded from StockCheck.com
I N T R O D U C T I O N A N D S P E C I F I C A T I O N S
1
Specifications
Right Display (Wavelength)
Units
Range
Resolution
Bargraph
Brightness
General Specifications
Operating Temperature
Storage Temperature
Humidity (non-condensing)
Overall Dimensions
Weight
Input Port Diameter
Input Type
Optical Stand Accommodations
Fiber Optic Adapter
Connector to Main Unit
1. Add + 0.5% from 400 to 450 nm and from 990 to 1100 nm.
2. Minimum optical power required for accurate wavelength measurement:
400 nm ~ 500 nm
100 uW (-10 dBm)
500 nm ~ 1100 nm
10 uW (-20 dBm)
For power measurement below these levels, use MANUAL wavelength mode.
3. For input power > 100 mW:
Add +0.05% / 100 mW to power accuracy
Add +0.05 nm / 100 mW to wavelength accuracy
Example: 500 mW, 850 nm, accuracy = +4.2%, +1.2 nm
4. MANUAL wavelength mode: Add +0.5% for automatic wavelength correction.
5. Wavelength accuracy between 450 nm to 1100 nm.
Our goal is to design and produce the best optical test equipment available
anywhere. To achieve this, we need your ideas and comments on ways we can
improve out products. We invite you to contact us at any time with your
suggestions.
-1
Ä
nm, cm
,
500.0 to 1000.0 nm, 20,000 to 10,000 cm
-1
.1 nm, .1 cm
Right Display (Wavelength)
Percent of full scale power
5 settings
o
o
+ 15
C to +35
C
o
o
-40
C to +70
C
> 100 W: <85% RH
< 100 W: <70% RH
68.6 mm diameter x 30.1 mm
< 0.5 kg
6 mm
Integrating sphere
8/32 tapped hole centered
FC, ST, Bare Fiber
26 pin high-density sub-D