Tablet and external interface for matrix switch. (20 pages)
Summary of Contents for FSN FM-C5501DV
Page 1
4K UHD Monitor Instructions for Use FM-C5501DV FM-A5504DS FM-A5504DG FM-D5801DV FM-D5802DV Before connecting, operating or adjusting this product, please read this instruction booklet carefully and completely. English...
Product Description / Intended Use This product from FSN Medical Technologies is a high-end surgical display monitor designed for advanced digital OR applications. This medical display is uniquely equipped to handle tasks in the demanding operating room environment. Performance characteristics include: •...
Symbol Defi nitions The following symbols appear on the product, its labeling, or the product packing. Each symbol carries a special defi nition, as defi ned below: Dangerous : High Consult accompanying Power adapter Voltage documents lndicates equipotential Direct current Unique Device Identifi er earth ground Indicates protective...
Warnings and Precautions Caution Information CAUTION RISK OF ELECTRICAL SHOCK DO NOT OPEN This symbol alerts the user that important literature concerning the operation of this unit has been included. Therefore, it should be read carefully in order to avoid potential problems. This symbol warns users that un-insulated voltage within the unit may have suffi cient magnitude to cause electrical shock.
Page 6
Use the proper power cord with correct attachment plug type. If the power source is 120 V AC, use a power cord which is a Hospital Grade Power Cord with NEMA 5-15 style plug, labeled for 125 volts AC with UL and C-UL approvals. If the power source is a 240 V AC supply, use the tan- dem (T blade) type attachment plug with ground conductor power cord that meets the respec- tive European country’s safety regulations.
Page 7
Safety Instructions On Safety 1. Before connecting the AC power cord to the DC adapter outlet make sure the voltage desig- nation of the DC adapter corresponds to the local electrical supply. 2. Never insert anything metallic into the cabinet openings of the medical LCD monitor. Doing so may create the danger of electric shock.
Page 8
Environmental Conditions for Operation and Storage Temperature range within 0°C to 40°C(operation), -20°C to 60°C (storage) Relative humidity range 10% to 85% Atmospheric pressure range within 500 to 1060hPa. On Installation 1. Openings in the medical LCD monitor cabinet are provided for ventilation. To prevent over- heating, these openings should not be blocked or covered.
Page 9
Returned Product After troubleshooting, if problems persist, disinfect the monitor and return it to FSN using the original packaging. Include the accessories that came with the monitor in the return shipment. Please enclose a brief explanation of the malfunction.
Electromagnetic Compatibility This medical monitor unit has been designed and tested to comply with IEC 60601-1-2:2014/ AMD1:2020 requirements for EMC with other devices. To ensure electromagnetic compatibility (EMC), the monitor must be installed and operated according to the EMC information provided in this Instructions for Use.
Page 11
1. Guidance and manufacturer’s declaration - electromagnetic emission The medical LCD monitor is intended for use in the electromagnetic environment specifi ed below. The user of the device should make sure that the medical LCD monitor is operated in such an environment. Interference emission measurements Conformity level Electromagnetic environment -guidance...
Page 12
3. For the use of ME devices in professional healthcare facilities. Test specifi cation for ENCLOSURE PORT IMMUNITY to RF wireless communica- tions equipment (according to IEC 60601-1-2:2014) The medical LCD monitor is intended for use in the electromagnetic environment specifi ed below. The user of the medical LCD monitor should make sure that it is used in such an environment.
Page 13
4. Guidance and manufacturer’s declaration – electromagnetic immunity – for equipment and systems that are not life-supporting The medical LCD monitor is intended for use in the electromagnetic environment specifi ed below. The user of the medical LCD monitor should make sure that it is used in such an environment. Interference IEC 60601-1- Conformity...
Page 14
5. Recommended separation distances between portable and mobile RF communications equipment and the medical LCD monitor The medical LCD monitor is intended for use in the electromagnetic environment in which the RF distur- bances are controlled. The user of the device can help prevent electromagnetic interference by maintai- ning a minimum distance between portable and mobile RF communications equipment (transmitters) and the device –...
FM-C5501DV 82.5 / 3.25 1278 mm / 50.31 35.4 1.39 29.76 600 / 23.62 15.75 ON/OFF DISPLAY SINGLE LINK GROUND INLET SWITCH DVI 2 DVI 1 (SVC) PORT 2 IN DVI IN DVI OUT 5V OUT 2 DUAL LINK IN...
Page 17
FM-A5504DS 93.5 / 3.68 1292.5 mm / 50.89 57.5 2.26 1278 mm / 50.31 770.5 29.76 30.33 600 / 23.62 15.75 ON/OFF SDI 3 DVI 4 SDI 1 DVI 2 DISPLAY GROUND INLET SWITCH (SVC) PORT 2 IN 5V OUT 2 SDI 4 DVI 3 SDI 2...
Page 18
FM-D5801DV, FM-D5802DV 1358 mm / 53.46 in. 105 / 4.13 60.8 1272.6 mm / 50.1 in. 2.39 32.36 28.5 600 / 23.62 15.75 ON/OFF DISPLAY SINGLE LINK GROUND INLET SWITCH (SVC) PORT 2 IN DVI 2 DVI 1 DVI IN DVI OUT 5V OUT 2 DUAL LINK IN RS232C...
Controls On Screen Display (OSD) When the OSD menu is When the OSD menu is Press to enable PIP (Pic- Press to turn power on/ activated, press to de- activated, press to move ture in Picture) function. off to the display’s front crease the adjustment of the menu selection LCD screen.
Page 20
On Screen Display (OSD) Menus FSN display monitors come equipped with a rich set of features for system set-up, image adjustments, and screen layout control. These features are managed through the On Screen Display, or OSD. Some options presented in the OSD are contextual and vary depending on the active input signal.
Page 21
On Screen Display (OSD) Menus FM-A5504DS, FM-A5504DG Submenus under the COLOR menu 1. GAMMA Select the appropriate gamma. (BYPASS,1.8, 2.0, 2.2, 2.4, 2.6, DICOM). 2. COLOR SPACE Select the color space setting. (NATIVE, sRGB, BT.2020, or AUTO) 3. COLOR MODE Changes the image color setting. (C1, C2, C3, USER) 4.
Page 22
4. OVER SCAN Adjusts the displayed size. (0~6) 5. DUAL DVI SOURCE or SDI SOURCE SELECT FM-C5501DV, FM-D5801DV: Select Dual-link DVI input source. (DVI-1, DVI-2, Dual) Note: SDI-QUAD or SDI-2SI will support 4K quadrant and 2-sample interleave division when using SDI-1 through SDI-4 input signal connectors.
Page 23
On Screen Display (OSD) Menus FM-C5501DV, FM-D5801DV, FM-D5802DV Submenus under the OSD SETTING menu 1. LANGUAGE Changes the OSD language. (10 languages) 2. OSD TRANS Adjusts the OSD transparency. 3. OSD POSITION Changes the OSD position. (9 Positions) 4. OSD MENU TIME Adjusts the length of time the OSD Menu is present on the screen. (range: 10~60 seconds) 5.
• 3840 x 2160 @60Hz 135.00 60.00 594.00 DVI dual-link mode only. DVI dual-link mode is only on FM-C5501DV, FM-D5801DV, FM-D5802DV. 12G SDI input port, quadrant and 2 sample interleave mode only. 12G SDI mode is only on FM-A5504DG. 8/2021...
Specifi cation FM-C5501DV Item Description Panel 55 inch TFT LCD (LED) Resolution 3840 x 2160 pixels Aspect Ratio 16 : 9 Active Area 1209.6 (H)mm x 680.4 (V)mm Pixel Pitch (mm) 0.315 x 0.315 Response Time (typical) 6 ms (gray to gray) Number of Colors 1.08 Billion...
Page 28
Specifi cation FM-A5504DS, FM-A5504DG Item Description Panel 55 inch TFT LCD (LED) Resolution 3840 x 2160 pixels Aspect Ratio 16 : 9 Active Area 1209.6 (H)mm x 680.4 (V)mm Pixel Pitch (mm) 0.315 x 0.315 Response Time (typical) 8 ms (gray to gray) Number of Colors 1.07 Billion Brightness (typical)
Page 29
Specifi cation FM-D5801DV, FM-D5802DV Item Description Panel 58 inch TFT LCD (LED) Resolution 3840 x 2160 pixels Aspect Ratio 16 : 9 Active Area 1270.08 (H)mm x 721.44 (V)mm Pixel Pitch (mm) 0.33075 x 0.334 Response Time (typical) 9.5 ms (gray to gray) Number of Colors 1.07 Billion 700 cd/m 2...
Cleaning Instructions Follow your hospital protocol for the handling of blood and body fl uids. Clean the display with a diluted mixture of mild detergent and water. Use a soft cotton towel or swab. Use of certain detergents may cause degradation to the labels and plastic components of the product. Consult cleanser manufacturer to see if agent is compatible.
Page 32
Thank you for choosing our product. Service Contact the appropriate customer service listed below for product information or assistance. Warranty One year, parts and labor. EC Representative KTR Europe GmbH Mergenthalerallee 77, Eschborn 65760, Germany Tel : +49(0)6196-887170 FORESEESON GmbH Industriestrasse 38a, 63150 Heusenstamm, Germany Tel.
Need help?
Do you have a question about the FM-C5501DV and is the answer not in the manual?