10D1475 MINI-MAG MAGNETIC FLOWMETER INSTRUCTION BULLETIN
1.4 Specifications
Power Requirements
Power Consumption
Flowmeter Characteristics
Meter Size/Flow Capacity
Any maximum (range) between 0.02 x Meter Capacity & 1.0 x Me-
ter Capacity is available with HART or standard software.
Meter Size
inch
mm
⁄
1
3
10
⁄
5
4
32
⁄
1
6
4
⁄
3
10
8
⁄
1
15
2
1
25
⁄
1
1
40
2
2
50
3
80
4
100
* Flow Velocity (ft/s) = (Operating GPM x 32.81)/METER CAPACITY
Span
Rangeability
Minimum Liquid Conductivity
System Accuracy
(At Reference Conditions)
Frequency Output
RFI Protection
1-8
See Section 1.3 Model Number Breakdown
Refer to Signal Converter Instruction Bulletin
NOTE
TABLE 1-2. METER CAPACITY
METER
CAPACITY*
gpm
gpm
1.06
0.021
2.12
0.042
5.28
0.11
11.9
0.24
26.4
0.53
52.8
1.06
158.5
3.17
264.2
5.28
792.5
15.85
1057
21.14
Factory set at specified range between extremes
listed in table above; can be field adjusted.
50:1
5 µS/cm
Flowrate
< 2% Meter Capacity
≥ 2% Meter Capacity
Class 3-abc-0.5% (30 V/m-27 to 446 MHz) Per
SAMA Standard PMC 33.1-1978.
Flow Ranges
0 to value tabulated
Minimum
lpm
0.08
0.16
0.40
0.90
2.00
4.00
12.0
3
m
/h
1.2
3.6
4.8
Accuracy
± 0.01% Meter Capacity
± 0.5% of rate
Maximum
gpm
lpm
1.06
4.0
2.12
8.0
5.28
20
11.9
45
26.4
100
52.8
200
158
600
3
m
/h
264
60
792
180
1057
240
INTRODUCTION