• Press and hold "Alarm Silence", "Soft" &
Calibration
(Control Unit must be
in Stand-By mode and
• Unit will display calibration instructions.
the mattress hose
disconnected
)
• Place new Fowler Transmitter (in bed or
Auto Fowler Set-up
(Control Unit must be
• Press and hold "A/P" & "Fowler" keys
in Stand-By
mode)
• Tilt transmitter/bed to fowler position, unit
• Routine will auto time out in 2 minutes if
• During Alternating Pressure therapy,
Set A/P Low Value
K-1 Elite
(Control Unit
Must be Powered on)
• Use the (L) "+" and "-" keys to adjust the
K-1oem
(Control Unit must be
• Press the AP key to save setting and
in Stand-By
mode)
• Press and hold "AP" & "LAL" keys until
Calibration
(Control Unit must be
• Unit will display calibration instructions.
in Stand-By mode and
the mattress hose
disconnected
)
• Place new Fowler Transmitter (in bed or
Auto Fowler Set-up
(Control Unit must be
• Press and hold "Patient Setup" & "Mode"
in Stand-By
mode)
• Unit will display set up instructions.
• Routine will auto time out in 2 minutes if
• During Alternating Pressure therapy,
Set A/P Low Value
K-2 Elite
(Control Unit
• Use the (L) "+" and "-" knob to adjust the
Must be Powered on)
K-2oem
(Control Unit must be
in Stand-By
mode)
• Press AP key to save setting and

CONTROL UNIT CALIBRATION AND SETTINGS

K-1 ELITE & K-1 OEM SERIES
K-1 ELITE
"Firm" keys until unit beeps. (Approx. 3
seconds).
external) to the flat position.
until unit beeps. (Approx. 3 seconds).
beeps, unit returns to stand-by mode
no transmitter is detected.
Press and hold (about 3 seconds) the
"AP" key. AP settings will be displayed on
the screen.
Low AP percentage value (0% to 80%
mmHg of high pressure setting).
continue AP therapy. Unit will time out
and setting will be set automatically
within 30 seconds.
K-2 ELITE & K-2 OEM SERIES
K-2 ELITE
unit beeps. (Approx. 3 seconds).
external) to the flat position.
keys until unit beeps. (Approx. 3
seconds).
no transmitter is detected.
Press the "Mode" key. AP settings will be
displayed on the screen.
Low AP percentage value (0% to 75%
mmHg of high pressure setting). Max
can be press to activate ZIP (Zero
Interface Pressure, 0% mmHg of high
pressure setting).
continue AP therapy.
K-1 & K-2 Series Owner's Manual
K-1oem
• Press and hold "Mode", "Alarm Silence" & "Lock" keys until
unit beeps. (Approx. 3 seconds). 7-segment will display "C"
for calibration, Static LEDs will be flashing.
• Press "Mode" key to complete calibration and end routine.
• Place new Fowler Transmitter (in bed or external) to the flat
position.
• Press and hold "Fowler" & "Max" keys until unit beeps.
(Approx. 3 seconds). "L" will be shown on the 7- segment
display.
• Tilt transmitter/bed to fowler position, unit beeps and returns
to stand-by mode.
• To end routine without new fowler, press "Fowler" key.
• Routine will auto time out in 30 seconds if no transmitter is
detected.
• Press and hold "Alarm Silence" & "Fowler" keys until unit
beeps. (Approx. 3 seconds).
• 7- segment will display (AP Low modes)
"0" (0% mmHg of high comfort pressure setting),
"2" (25% mmHg of high comfort pressure setting), "5" (50%
mmHg of high comfort pressure setting) or "7" (75% mmHg of
high comfort pressure setting)
• Use the Soft /Firm keys to toggle between AP Low modes
• To end routine and accept setting, press "Power" key.
Routine will auto time out
K-2oem
• Press and hold "Mode", "Time" & "Alarm Silence" keys until
unit beeps. (Approx. 3 seconds). Static, AP, Fowler LED will
turn and AP LED will be flashing.
Press Mode key to complete calibration and end routine.
• Place new Fowler Transmitter (in bed or external) to the flat
position.
• Press and hold "Mode" & "Lock" keys until unit beeps.
(Approx. 3 seconds). "L" will be shown on the 7-segment
display.
• Tilt transmitter/bed to fowler position, unit beeps, all LEDs
light up and unit returns to stand-by mode
• To end routine without new fowler, press "Fowler" key.
• Routine will auto time out in 60 seconds if no transmitter is
detected.
• Press and hold "Lock" & "Select" keys until unit beeps.
(Approx. 3 seconds).
• 7- segment will display (AP Low modes).
"0" (0% mmHg of high comfort pressure setting) or
"5" (50% mmHg of high comfort pressure setting)
• Use the Soft and Firm keys to toggle between AP Low modes
• To end routine and accept setting, press "Power" key.
Routine will auto time out in 30 seconds if no key is pressed.
Page 28