Make sure the CPR valve is in closed position, bolsters and air pad plugs are connected and locked in place before
inflating the mattress.
Make sure the left and the right side safety air bolsters (if available) as well as the bottom safety air pad are fully inflated
and up to the correct operating pressure and firm prior to placing a patient on the mattress.
Note: Not all functions are available on all models. {Display labels – page 12 & Figure 1- Page 16}
Note: Weight scale on overlay display for models K-1oem ALT-AIRE and K-2oem ALT-AIRE + is for reference
only; the comfort level on the mattress system can be adjusted for patient's comfort by pressing the
FIRM key accordingly. Please refer to "PATIENT COMFORT CONTROL LEVEL" on page 19 for additional information.
INITIAL POWER UP
(K-1 ELITE, K-2 ELITE) During initial power up when power cord (Q) is plugged into the power source (X), the control
unit will display "KAP MEDICAL, K-1 APP SYSTEM" on K-1 system and "KAP MEDICAL, KARE-ZONE, DYNAMIC LAL
SYSTEM" on K-2 system for a brief moment and then goes through a system initialization routine for a few seconds.
Once the routine is completed the display (D) will read "KAP MEDICAL, STANDBY".
(K-1oem, K-2oem) During initial power up when power cord (Q) is plugged into the power source (X), the control unit will
have an illuminated amber light indicating that the unit is in Standby mode.
POWER (SB)
Press Power key, the (amber) Standby LED will turn off and the (green) LED turns on. The control unit will turn on.
MAX INFLATE (W)
Press MAX Inflate (W) key, the (Max Inflate) LED will turn on (K-1, K-1oem, K-2). Press Select key (S)
Max Inflate LED
turns on (K-2oem). This mode is used to rapidly inflate the mattress. During this mode a series of
beeps will sound every 3 minutes as a reminder that MAX Inflate mode has been activated.
In Max Inflate, the entire mattress will be pressurized to 35 ± 5 mmHg. The mattress (B) will inflate to its normal size
within 5~15 minutes. MAX Inflate mode will deactivate after 30 minutes.
THERAPY (STATIC) (T)
To set Static mode press (T) Therapy
mode indicator light is illuminated (and/or the screen displays THERAPY on K-2 Elite model only). The unit will go into
"THERAPY" mode.
In THERAPY mode all the air cushions in the mattress will be maintained at the same pressure and can be adjusted to
the desired patient comfort control level.

OPERATING INSTRUCTIONS

key (K-1 Elite model only) or (M) Mode
K-1 & K-2 Series Owner's Manual
SOFT key or
until the
key until the Static
Page 18