LOW PRESSURE (LP)
In the event of hose disconnection or other problem that causes major air leakage, the Micro-controller will activate an
audio/visual signal to alert the caregiver by flashing "LOW PRESSURE" on the digital display (K-1 Elite and K-2 Elite
models) or flashing the amber "LOW PRESSURE" LED (K-1oem and K-2oem models) and turning on the buzzer. Once the
low-pressure problem is fixed, the audio/visual signal will cease and the unit will resume its operating set mode.
BOLSTERED MATTRESS (RSB)
The left and the right bolsters can be manually deflated by disconnecting the bolster deflate connector which is at the
bottom right corner (patient's right) of the mattress.
PLEASE NOTE: Before using the mattress please make sure that the bolster deflate connector is
re-connected back into the bolster deflate valve.
BOTTOM SAFETY AIR PAD (OPTIONAL)
The bottom safety air pad can be manually deflated by disconnecting the air pad deflate connector which is at the bottom
right corner (patient's right) of the mattress.
PLEASE NOTE: Before using the mattress please make sure that the air pad deflate connector is re-connected back
into the air pad deflate valve.
SAC Support Air Cushion (K-140) (OPTIONAL)
SAC (Optional) foot support air cushion can be used to support patient's feet and to provide zero interface pressures in
the heel area.
FOAM AIRE MATTRESS AS A NON-POWERED MATTRESS
Foam Aire mattress can be used as a non-powered mattress without the control unit. Before using the mattress connect the
control unit to the mattress and inflate the mattress by setting the control unit comfort level to 5 on K-1oem, K-2oem and
setting of 22mmHg on K-1 Elite, K-2 Elite. Once the control unit fills the mattress and the pressure stabilizes then turn the
unit off and disconnect the hose assembly from the mattress and control unit and store the control unit and the hose
assembly in a storage area.
FOAM AIRE MATTRESS AS A POWERED MATTRESS
If need arises the Foam Aire mattress can be converted into a powered mattress by simply connecting the control unit to the
mattress using the hose assembly provided with the system. For operating instructions in powered mode please refer to
OPERATING INSTRUCTIONS ABOVE.
RECOMMENDED PRESSURE SETTINGS (ALL MATTRESSES)
1. For rapid inflation of the mattress, press (W) "MAX Inflate" key (K-1 Elite, K-1oem and K-2 Elite) or (S) "Select" key
(K-2oem) until the Max Inflate LED turns on.
2. For extra firm support during patient ingress or egress, patient wound care, patient turning or patient cleaning it is
recommended to set the mattress pressure to maximum pressure by activating MAX Inflate mode.
& EVEN BOLSTERED MATTRESS (ESB)