Microtech AiDEX User Manual page 58

Continuous glucose monitoring system
Hide thumbs Also See for AiDEX:
Table of Contents

Advertisement

Specifications
Model number
Operating temperature
Operating humidity
Storage and transporta-
tion temperature
Storage and transporta-
tion humidity
Storage and transporta-
tion pressure
Ingress protection level
Use life
Detection range
Measurement accuracy
Wireless frequency and
bandwidth
Wireless modulation
Radiated power
54
Item
When the glucose concentration >4.2mmol/l
(75mg/dL), the accuracy deviation of the sensor
does not exceed ±20%; when the glucose con-
centration ≤4.2 (75mg/dL), the accuracy deviation
does not exceed ±1mmol/l (18mg/dL).
Frequency: 2.402GHz ~ 2.48 GHz
Bandwidth: 1Mbps
GFSK
-2dBm
Subcomponent
Transmitter
G7-T01
G7-T01A
G7-T01B
5-40℃(41-104 ℉)
10-93%(non-condensing)
-20℃ -60℃
5-95%(non-condensing)
700hpa~1060hpa
IPX7
G7-S01: 14 days
G7-S01A: 10 days
4 Years
G7-S01B: 7 days
Shelf life: 1 year
2.0mmol/L-25.0 mmol/L
Sensor
G7-S01
G7-S01A
G7-P02
G7-S01B
4℃ -30℃
-20℃ -60℃
4 Years
PDA
IPX0

Hide quick links: