PDA alarms
Alarm de-
scription
Low Blood
Glucose
High Blood
Glucose
Sensor
Expired
Sensor Error
Other
Unspecified
Errors
40
Priority
Perform a finger-tip glucose measurement
using an approved glucose meter to confirm
the sensor glucose reading. If the measure-
Medium
ment indicates low blood glucose, please
take actions to raise blood glucose immedi-
ately and call for medical assistance
Perform a finger-tip glucose measurement
using an approved glucose meter to confirm
the sensor glucose reading. If the measure-
Low
ment indicates high blood glucose, please
take medicine or inject appropriate amount
of insulin according to doctor's prescription.
Remove the current glucose sensor and
Low
replace it with a new sensor.
Check if the current sensor has peeled off
or partially peeled off from the skin. If yes,
please remove the current sensor and re-
place it with a new sensor. If the sensor is
Low
still firmly attached to the skin, but the Sensor
Error message continue to appear for more
than 30 minutes, Please remove the current
sensor and replace it with a new sensor.
Please switch off the PDA and restart it. If the
Low
error message continue to appear, please
contact our customer service.
Solution/Action