• When used in a medical facility, the operator should
wear gloves while using.
Blood glucose test module
precautions
• This function should only be used in vitro and can
only be used with MicroTech's Exactive EQ blood
glucose test strips. Use of other brand test strips will
result in incorrect test results.
• The blood glucose test function can only be used to
determine blood glucose levels using whole blood
samples. Do not use serum or plasma samples.
• The blood glucose test function is not intended for
neonatal applications.
• Test results may not be accurate for blood hematocrit
of above 55% or below 30%.
• Blood containing high levels of vitamin C or other
reducing agents can lead to inaccurate results.
• The blood glucose meter test range is 1.1-33.3mmol/
6