LaMotte DC1500 Manual page 10

Ammonia colorimeter kit
Hide thumbs Also See for DC1500:
Table of Contents

Advertisement

3. Scroll to and select 008
4. Rinse a clean tube (0290)
5. Insert tube into chamber, close
6. Remove tube from colorimeter.
7.
1.800.561.8187
Ammonia-N HR from menu.
three times with the ammonia
nitrogen standard. Fill to the
10 mL line with the ammonia
nitrogen standard. Cap the tube.
Dry the tube with a lint-free
cloth.
lid and select Scan Blank.
Add 8 drops of Ammonia
Nitrogen Reagent #1 (V-4797).
Cap and mix.
Use 1.0 mL pipet (0354) to add
1.0 mL of *Ammonia Nitrogen
Reagent #2 (V-4798).
www.
008 Ammonia-N HR
Scan Blank
Scan Sample
12:00:00
001/500