DC1500
AMMONIA
ANALYSIS
PROCEDURE
1.
Fill Water Sample Collecting Bottle (0688) with sample water.
2.
Press and hold
3.
Press
ENTER
4.
Scroll to and select 008 Ammonia-N HR from menu.
5.
Rinse tube (0290) with sample water. Fill to the 10 mL line.
Cap and wipe dry.
6.
Insert tube. Close lid. Select Scan Blank.
7.
Remove tube. Add 8 drops of Ammonia Nitrogen Reagent #1.
Cap and mix.
8.
Use 1.0 mL pipet (0354) to add 1.0 mL of *Ammonia Nitrogen
Reagent #2.
9.
Cap and invert to mix. Wait 5 minutes for full color
development.
10. Insert tube into chamber. Close lid. Select Scan Sample.
Record result as AMMONIA-NITROGEN.
See manual for calculations.