Electrical Connections - VERDER CARBOLITE GERO E Series Installation, Operation And Maintenance Instructions

1200 c tube furnaces
Table of Contents

Advertisement

2.0   Installation
2.4

Electrical Connections

Connection by a qualified electrician is recommended.
All furnaces in the range are single-phase and operate over the voltage ranges 100 to
110 volts or 200 - 240 volts. Check the rating label before connection, see "Warning
Labels" for location of rating label. To check that the OP.Hi setting is correct for the
appropriate voltage, or to change to another setting, follow the instructions in section
0.1.
The electrical supply should be fused at the next standard size equal to or higher than
the design current. Where a supply cable is supplied there are internal supply fuses, in
which case customer fusing is preferred but not essential. See section 10.0 for furnace
rating power, current and fusing information.
The external connection should be either a permanent connection to a fused isolator
supply or to a socket on a fused isolator supply; the isolating switch must operate on
both conductors. Make sure that the isolator switch is within easy reach of the furnace
operator.
Terminal
Supply
Label
L
N
1-phase
PE
16
Terminal Connection Details
UK/ Europe
Cable Colour
Brown
Blue
Green/ Yellow
USA/
Canada
Common
Live -
Cable
Neutral
Colours
Red or
To live
Black
White
To neutral
Green or
To earth
Green/
(ground)
Yellow
Supply Type
Reversible or
Live - Live
to either power
conductor
to the other power
conductor
to earth (ground)