Read and then remove the sticker (see Figure ) from the instrument calibration
cover. Checking and setting of the alarm point(s) is discussed in detail below and
in the sections Setting Alarm-Points and Calibration Controls in this manual.
Please set the alarm point(s) on this insturment
to conform to your requirements. The
factory-set alarm points may be incorrect for
your use.
Refer to the instrument manual for more
information on setting alarm points.
FAILURE TO RESET THE ALARM POINT(S)
MAY RESULT IN EXCESSIVE ALARMS OR
LACK OF SENSITIVITY.
Initial power-up (by way of the power switch on the left side panel) will
momentarily activate the internal front-panel lights, sound the audio, and display
"8888" on the 4-digit LED display. The firmware version number (39665Nyy) is
then displayed as "396" and "65yy" (where yy represents the current version
number).
When the instrument has finished measuring background, it will display the
current radiation reading and begin checking for an alarm condition.

Radiation Units

The Product may be calibrated for almost any desired radiation units of measure.
Common units of measure include mR/hr, µR/hr, R/hr, mSv/h, µSv/h, cps, cpm,
and kcpm. In each case, the unit of measure is indicated underneath the four-
digit display. Throughout the rest of this manual, the notation <units> will be used
as a substitute.
Figure 3. Alarm Point Notice
Digital Wall Mount-Area Monitor
Getting Started
huk02.eps
7