The following steps describe how to properly install the cables for an OCIA StratosH application. The scanner must then be
configured to match the host's OCIA parameters. Cable installation alone does not guarantee that the StratosH will
communicate properly with the host system.
Configuration bar codes are located in the MetroSelect Configuration Guide (
MS2xxx Stratos Series Configuration Addendum (
1.
Turn off the host system.
2.
Plug the 10-pin RJ45 end of the OCIA interface cable (
OCIA to Host on the bottom of the StratosH. Refer to figure on page 25.
3.
Connect the other end of the OCIA Interface cable (
host device.
Before continuing verify that the OCIA cable pinouts match the host system's and is connected to the appropriate
interface socket on the scanner. An incorrect cable connection can cause communication problems or potential
damage to the scanner and/or terminal.
Manufacturers Note: Plugging the scanner into the serial port of the host device does not guarantee that scanned
information will appear at the host device. A software driver and correct configuration settings are also required for
proper communication to occur.
4.
Plug the external power supply (
the StratosH.
xxx*
Specifies international connection. See Optional Accessories in the Introduction section of this guide for a complete listing.
Check the AC input requirements of the power supply to make sure the voltage matches the AC outlet. The outlet
should be located near the equipment and be easily accessible.
Metrologic recommends using a switched AC outlet. The switch should be located on the operator's side of the
checkstand in close proximity to the StratosH to facilitate calibration and service of the unit.
5.
Connect AC power to the transformer. If the AC outlet is equipped with an on/off switch, turn the power on.
6.
Turn on the host system.
7.
Configure the StratosH to match the host system's OCIA parameters.
Refer to the MetroSelect Configuration Guide (
Setting OCIA Parameters.
Caution: To maintain compliance with applicable standards, all circuits connected to the scanner must meet the requirements for SELV
(Safety Extra Low Voltage) according to EN/IEC 60950.
To maintain compliance with standard CSA C22.2 No. 60950/UL 60950 and norm EN/IEC 60950, the power source should
meet applicable performance requirements for a limited power source.
24
S
)
NTERFACE
PECIFIC
46-46
xxx*
MLPN
00-02034x).
MLPN
57-57015x-N-3) into the 10-pin socket labeled Scanner
MLPN
57-57015x-N-3) to the appropriate communication port on the
MLPN
) into the 3-pin Molex socket labeled, DC Power In, on the bottom of
00-02407x) under Section I: OCIA for Enabling and
MLPN
00-02407x) and the
MLPN