7.1.2 Do not mix with dangerous goods during transportation.
7.1.3 Avoid sunlight, rain or snow during transportation
7.2 Storage
7.2.1 Do not mix with toxic, corrosive, flammable and explosive materials during storage.
7.2.2 The product should be stored in an environment where the relative humidity does not exceed 10% - 93%,
the atmospheric pressure is 70kPa - 106kPa, and the temperature is -20°C - +40°C.
7.3 Maintenance
7.3.1 This equipment should be handled with care, far away from the source of the earthquake, and should be
installed or stored in a cool, dry and ventilated place.
7.3.2 When the device is not in use, turn off the power switch and unplug the power plug; when not in use for a
long time, charge the main unit every 3 months.
7.3.3 To clean or disinfect, wipe the surface with clean water or disinfectant, do not soak.
7.3.4 After use, check if there is any resin remaining on the surface of lamp, so as not to affect the life of the
front joint or affect the curing effect.

8. Troubleshooting

Faulty
No reaction
Insufficient Light intensity
Not charging when con-
necting with the adapter
Less working life after
fully charge
The mode indicator twin-
kles when charging
If the fault still cannot be resolved, please contact your local dealer or our company.

9. Electromagnetic compatibility

Attention:
●This Curing Light meets the relevant requirements of YY 0505 standard electromagnetic compatibility. The basic
performance of the light curing machine is: the front LED light can emit blue (UV) when working.
●The user should install and use it according to the electromagnetic compatibility information provided by the
accompanying documents.
●Portable and mobile radio frequency communication equipment may affect the performance of the curing light,
avoid strong electromagnetic interference when using it, such as close to mobile phones, microwave ovens, etc.;
●The guidelines and manufacturer's declaration are detailed in the attachment.
●Without the express consent of Guilin Refine Medical Instrument Co., Ltd., unauthorized changes or
modifications to the equipment may cause damage to the equipment or other
The electromagnetic compatibility of the equipment.
Alert:
●This light curing machine should not be used close to or stacked with other equipment. If it must be used close
or stacked, it should be observed and verified that it can be used in the configuration used.
Normal operation.
● Except for the cables sold by the manufacturer of the curing light as spare parts of internal components, the use
of accessories and cables other than those specified may cause the curing light
Increase in emission or decrease in immunity.
● The following cables must be used to comply with electromagnetic emission and anti-interference requirements:
Cable Name
Power Line (Output end)
● Key components of electromagnetic compatibility
Possible cause
1.Out of battery
2.Faulty of battery
3.Battery protection
There is resin remaining on the
surface of the transparent cover
of light guide.
1.The adapter is not connected
well
2.Faulty of adapter or incompat-
ible
3.The charging point is impurity.
Capacity of battery decreased
Low voltage
Cable Type
Unshielded Parallel Lines
1.Charge/Change a new battery
2.Change a new battery
3.Activate it by putting it on the
base
1.Remove the resin
2.Change the transparent cover
1.Reconnect.
2.Change the adapter
3.Clean the inner connector of
charging base by the alcohol
Change a new battery
Back to normal after 15-minute
charging.
Cable Length
1.2 m
5
solutions