Operation - Swann CURING LIGHT Instruction Manual

Curing light
Table of Contents

Advertisement

Modes and time setting
Effective area of irradiation
light
3.2 Environmental factors
°C

3.2.1 Operation: +5

3.2.2 Relative humidity: 10% - 93%
3.2.3 Atmospheric pressure: 70kPa-106kPa
3.3 Equipment safety classification
3.3.1 Classified by type of protection against electric shock:
3.3.2 Classified by degree of protection against electric shock: Type B application part
3.3.3 Classified according to the degree of protection against incoming liquid: Ordinary equipment (IPXO), non-
waterproof
3.3.4 Classified by operation mode: Short-term operation equipment
3.3.5 Classification of safety degree when using flammable anesthetic gas mixed with air or oxygen or nitrous
oxide: Equipment that cannot be used with flammable anesthetic gas mixed with air or oxygen or nitrous oxide.
4 Installation and demounting
Insert the large end of the light guide (front connector) into the mounting hole at the upper end of the curing light
(please insert it into the bottom in a rotating way, do not be obliquely).
Clamp and install the light hood at the root of the light guide (front joint).
When removing or replacing the light guide, proceed in the opposite direction above.
5 Operation
5.1 When the machine is under stand-by mode, press the time/mode key to light up the screen.
5.2 Mode setting: press the time/mode key for more than 2 seconds to select the mode; during mode switching,
the display screen has a transition picture (the picture shows the full English name of the mode to be switched to),
and every time the key is pressed, there's a beep. Without operating within 1 minute, the screen will turn off and
the main unit enters the sleep state.
5.3 Time setting: press the time/mode key to select time, see table 1 for settable curing time.
5.4 Point the light guide rod (front connector) light source at the right position, press the switch button and there's
a beep. The curing light will start to work as the mode setting. The display screen starts counting down from the
selected time, and after the timing ends, the display screen displays the setting time.
5.5 Before the end of setting time, at any time pressing the on/off key can finish the work.
5.6 After the end of an irradiation cycle, you can immediately press the switch to start the next irradiation cycle. If
the handle starts to become hot, please stop working until the handle is completely cooled down. It is recommend-
ed that the number of consecutive irradiations should not exceed 10 times.
5.7 If the transparent cover is stained with resin after use, please wipe it with a cotton cloth to avoid affecting the
light intensity.
5.8 The effective light intensity of this machine is many times higher than that of the halogen lamp curing light, and
the curing depth of the light-curing composite resin is not less than 4mm for 10 seconds.
5.9 The front connector at the front end can be removed, but high temperature and high-pressure sterilization is
not allowed, otherwise it will cause damage.
6 Low battery prompt and charging
6.1 Low voltage indication: The battery symbol on the display starts to flicker, accompanied by a sound prompt;
6.2 When charging, insert the charging cable into the TYPE-C socket at the back of the base, the blue light at the
bottom of the base will light up, and the brightness will be changing;
6.3 Correctly put the main unit into the charging base, the base display will automatically light up and display the
power;
6.4 The charging base has a light meter function. When the base is in standby mode, the display is off. Only when
the main unit is charged or when testing light intensity that the display will automatically light up.If it is not charged
or measured, the base display will automatically turn off.
7 Transportation, storage and maintenance
7.1 Transportation
7.1.1 During transportation, avoid shocking and vibration, handle with care and avoid inversion.
TURBO mode: Blue and purple light, 2500mW/cm2 for 1s, 3s
NORMAL mode: Blue and purple light, 1200mW/cm2 for 5s, 10s, 15s, 20s
ORTHO mode: Blue and purple light, 2500mW/cm2 for 3s to work 10 times
every cycle, or 5s to work 10 times every cycle, interval 1s
CHECK mode: purple light, 900mW/cm2 for 30s, 60s
66.4mm²
°C
- +40
Class Ⅱ
4