Configuration Mode - Dräger Babytherm 8004 Operating Instructions Manual

Infant warmer system
Hide thumbs Also See for Babytherm 8004:
Table of Contents

Advertisement

Operation

Configuration Mode

Configuration Mode
Possible Activities
Code
Activity
c01
Displaying current software version
c02
Selecting skin temperature alarm threshold
c03
Adjusting speaker volume for alarms
c04
Hours of operation, phototherapy
c05
Testing control panel keys
c06
Testing nurse call
c07
Selecting language
c08
Adjusting display contrast
Err
Error reporting
NOTE: The control panel keys and displays have a
different purpose/meaning than in radiant heater
operation mode!
1 Activate/cancel configuration mode
2 Information regarding current activity
3 Abbreviation of current activity
4 Code of current activity
5 Advance to / back to activity
6 Increase/decrease values for activity selected
7 Activate/leave error reporting mode
8 State changes within activities
9 Display setting for current activity
Activating configuration mode
1 Keep »OK«-button pressed for 3 seconds.
A brief alert sound is generated.
2 Display message:
Configuration mode
Press OK-button to start.
1 Press OK button again. A brief alert sound is
generated. Configuration mode is now active.
The device defaults to activity »c01« and displays the
current software version
Displaying the software version
1 Display message:
Configuration mode
Mode c01: Software version.
2 The current software version is displayed, e.g.: 1.00.
3 The current activity code c01 is displayed.
44
2
9
10
3
5
4
1
skin °C
manual
mattress.°C
7
skin
man.
5
6
2
10
5
1
skin °C
manual
mattress.°C
skin
man.
1
10
2
5
3
1
skin °C
manual
mattress.°C
skin
man.
Babytherm 8010
15 min
°C
8
< 36°C
1
Babytherm 8010
15 min
°C
< 36°C
1
Babytherm 8010
15 min
°C
< 36°C

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents