Operating Concept

D
Status LEDs
Green LED continuously on during operation.
Red LED continuously on during power outages.
»OK« key for confirming settings and to acknowledge
alarms.
NOTE: If the »OK« key is not pressed within
10 seconds after changing a setting, the previous
settings will remain in effect!
Red indicator LED for "Inop." alarm.
Key for testing LEDs, displays, enunciator, and
audible alarm.
Key for silencing audible alarms.
Visual display of alarms:
Warning = red LED, Caution = yellow LED.
Text display, shows user messages and alerts.
Babytherm 8010 only – display and keypad for
operating mattress heater:
top: display of measured mattress temperature;
bottom: display of set value for mattress temperature
with keys for setting;
to the right: indicator for operation in extended range
high (>38 °C), low (<36 °C).
10
°C
5
1
Skin °C
manual
Skin
man.
Babytherm 8010
°C
Check
> 38°C
< 36°C
Mattress °C
OK
Display and keypad for radiant heater:
top: display for measured values of skin temperatures
close to core (tummy) and at periphery (toe);
bottom: Set value for skin temperature using keys for
changing set value and key for switching between
manual and skin temperature control mode. The
currently active mode is indicated by the respective
yellow LED;
Further to the right: bar graph for displaying heater
levels 1 through 10 for manual mode operation
Heater levels 1 to 3 = green bar elements,
Heater levels 4 to 10 = yellow bar elements,
and below, keys for setting heater levels.
On/Off keys for switching:
work light,
night light,
phototherapy (available option)
radiant heater.
The respective yellow LEDs are lit during operation.
Theory of Operation
Operating Concept
Inop.
17

Advertisement

Table of Contents
loading

This manual is also suitable for:

Babytherm 8010

Table of Contents