Attaching Ventilator Circuit Support Arm

Insert flexible support arm into hole in the left side of
head panel and tighten with knurled screw.

Supplies Storage Options

Units with height adjustable column:
– swivel cabinet (available option) with 2 compartments
and 1 swivel tray
Units without electric height adjustment:
– 2 open compartments in the column (standard),
or
– 1 swivel cabinet with 2 compartments and
1 swivel tray (available option).

Adjusting the Height (Option)

1 Briefly depress right pedal.
The bassinet is raised.
2 Briefly depress left pedal.
The bassinet is lowered.
Adjust to comfortable working height.
CAUTION !
Do not place any objects above or underneath
bassinet where they might be damaged when
height adjustments are made.
CAUTION !
Always use circuits and patient lines with sufficient
length for secure connection at minimum and
maximum height.
CAUTION !
Always observe the load limit of 50 kg on the
electric height adjustment.