Display for inactive measured value recording Display for date Pin 1: Pin 4: GND (reference Display for time Supply sensor potential) Mode selection button (output) Pin 5: n.c. (not connected) Data logger for connecting to external sensors DL200X operating manual for external sensors...
The data logger can now be configured via the software. Further detailed information about using the software is provided in the software manual which you can open from the help function of the Smart- Graph software. DL200X operating manual for external sensors...
USB power supply. Measured value detection and data recording are active; but measured value In the optionally available PoE design, the data logger can have its power display is inactive. supplied directly over the network connection. DL200X operating manual for external sensors...
Briefly pressing the mode selection button closes the message. It is only possible to configure if the data logger is not in logging mode (M3 or M4)! Tip: Note the BUS ID for measuring in the future directly on the sensor. DL200X operating manual for external sensors...
The sensor current is converted into voltage in the DL200X at a load. The load selection button to navigate to the configuration mode M7x. The option for re- has a resistance of <50 Ω. Both the sensor and the DL200X can be supplied via moving the first BUS sensor is shown on the display.
A0 / A1 Example LCD displays of sensor specification for registering Important: The DL200X is able to compensate cable resistances up an analogue sensor at the first input channel (AN0): to 10 Ω. The compensation typically reduces the influence of the cable resistance by a factor of 100.
0.5 µm, particles 1 µm, particles 5 µm, particles 10 µm, particles x µm, particles y µm, particles z µm, unknown, in- sensor types: active, wet-bulb temperature, global radiation, daylight, amount of rain, wind speed, wind direction, analogue/digital input, ground moisture, leaf wetness DL200X operating manual for external sensors...
Page 10
Temperature measurement with thermal element J type Temperature °C – Temperature measurement with thermal element J type Temperature °F – Temperature measurement with thermal element S type Temperature °C – Temperature measurement with thermal element S type Temperature °F – DL200X operating manual for external sensors...
Logging mode flashing static Reading the measured value memory Additionally, a hysteresis can be set by which the measured value must return to the valid range to switch off the alarm. DL200X operating manual for external sensors...
10.8. Cleaning on the display. Afterwards, the display returns to normal operation. The use of aggressive cleaning agents in the vicinity of the DL200X may damage Further detailed information about using the software is provided in the the humidity sensor beyond repair.
11. Technical data DL200X data logger for external sensors Measuring range 0 … 1 V Voltage input Accuracy ±200 µV ± 0.1 % of measured value 0 - 1 V Resolution < 500 µV Measuring range 2-cable operation: 4 … 20 mA, 3-cable operation: 0 … 20 mA Accuracy ±...
-0.48 mA … 0.00 mA Displayed: 0.00mA 0 … 20 mA Normal operation mode 0.00 mA … 20.00 mA Limit range 20.00 mA … 20.48 mA Displayed: 20.00mA Overcurrent approx. > 20.48 mA 0x52 (CHANNEL_OVERRANGE) DL200X operating manual for external sensors...
Page 15
Trotec GmbH Grebbener Straße 7 52525 Heinsberg +49 2452 962-400 +49 2452 962-200 info @ trotec.com www.trotec.com...