Calibration Manual
Remote Command
Remote Command
A
a
I
V
v
M
m
K
k
H
h
P
p
O
o
W
X
Y
T
t
C
F
R
r
u
d
<
>
0-9
-,.
<CR>
b
)
16
Table 2. Upper Display Remote Commands (725 with V2.xx and lower)
i
mA measurement
L
mA Loop Power
E
Voltage measurement
B
Single broadcast of the most recent upper display value and units
(
Single broadcast of most recent upper display value without header or units
Table 3. Lower Display Remote Commands (725)
mA measurement
mA source
mA 2W Sim
Voltage measurement
Voltage source
mV measurement
mV source
kHz measurement
kHz source
Hz measurement
Hz source
CPM measurement
CPM source
Ohms measurement (default 2W)
Select Ohms source
2-wire measurement (Ohms and RTDs)
3-wire measurement (Ohms and RTDs)
4-wire measurement (Ohms and RTDs)
Thermocouple measurement (default Type J). Use "S" command to select a
sensor type.
Thermocouple source (default Type J). Use "S" command to select a sensor type
Selects Centigrade ( T/C-RTD)
Selects Fahrenheit ( T/C-RTD)
RTD measurement mode (default Pt100 385). Use "S" command to select a
sensor type
RTD measurement mode (default Pt100 385). Use "S" command to select sensor
type.
Increment display source value
Decrement display source value
The < arrow key PC keyboard selects left arrow on 725
The > arrow key PC keyboard selects right arrow on 725
Enter a source value using ASCII characters 0,1,2,...9,-,.terminated by <CR>
(carriage return)
The 725 can receive a maximum of 10 characters prior to the carriage return.
Single Broadcast of most recent lower display value and units
Single broadcast of most recent lower display value without header or units.
Description